background image

 

8/10 

record. So for example if data record time interval is 30 minutes the graph shows time behaviour of radon 
concentration for last 64 hours.  
 

                                                         

 

 

 

Second  graph  shows  current  energy  spectrum  in  array  of  96  x  24  pixels  and  its  parameters.  Energy 

spectrum is divide into 96 interval and radon concentration is calculated from intervals d1-d2 (RaA) or from 
d1-d3 (RaA+RaC). Discriminations d1, d2, d3 are shows below graphs as well. 
 

Numbers of impulses in discrimination intervals are counted during spectra record time interval (Trec) 

or till then if one of particular intervals reaches 255 impulses. 

 

                       

 

 
 

Parameters on third data screen are especially service data but record time intervals are useful. 

 

 

Notice:    By  display  and  button  isnt  possible  to  setup  of  probe.  For  setting  of  probe  parameters  is 

neccessary to used the PC software application usbRADONview. 

7

 

Basic Maintenance 

 
Accumulator charging: 

   During portable use of radon probe is essential to monitor state of internal accumulator and recharge it if 
necessary. If accumulator is discharged  the probe automatically turns off.  The probe is switched on again 
powering  USB port. 
   Current state of accumulator can be determined in three ways: 

1) By LED diode ´STAT´ - If LED starts blinking in green-yellow color it indicates that system is working 
incorrectly and one of main case is low voltage of accumulator. (see paragraph “Operation Manual / 
LED diode “STAT””)  
2)  On  wireless  Central  Unit  display  –  Symbol  "#"  on  left  edge  of  display  second  line  means  that 
accumulator voltage is low. 
3) In usbRADONview application - where you can check current accumulator voltage. Voltage should 
not fall below 3.5 V, in limit conditions falls below 3.3V. 

Accumulator  is  charged  via  USB  port  using  supplied  USB  cable.  USB  cable  can  be  connected  to  PC  or  to 
supplied  power  adapter.  Connect  USB  cable  with  power  to  USB  port  of  probe.  LED  diode  ´CHRG´  next  to 
USB port of probe indicates charging status according to following chart: 

 

Summary of Contents for TSR3DNM

Page 1: ...ns Operation Manual v 1 2017 Table of Contents 1 Introduction 2 2 Description and Utilization 3 3 Scope of Delivery 4 4 Product Specification 5 5 Operating Instructions 6 6 Display Operation 6 7 Basic...

Page 2: ...asis of used product If you experience any problems with your product please contact us at TESLA Podebradska 56 186 180 66 Prague 9 www tesla cz 1 Introduction This document describes technical specif...

Page 3: ...The probe saves time records of these radon concentration values including values of humidity and temperature within its internal memory typically at an interval of 1 hour Next saved value is time re...

Page 4: ...ing results into internal memory The resulting values are downloaded after end of the measurement by B way B Probe connected via USB Using usbRADONview app and USB cable is possible to download result...

Page 5: ...ponse rate 30 minutes RaA 3 hours RaA RaC Radon records calculated from RaA quicker less sensitive calculated form RaA RaC slower more sensitive Measuring relative humidity range 10 90 Measuring tempe...

Page 6: ...accumulator is described in chapter Basic Maintenance Accumulator charging Power supply According to operation method the radon probe can be supplied 1 By internal accumulator for portable use Radon p...

Page 7: ...intro screen automatically after 5 seconds from intro screen directly from switched off display by pushing of button for longer than 1 second Changing of three data screens is cyclic without intro sc...

Page 8: ...cumulator charging During portable use of radon probe is essential to monitor state of internal accumulator and recharge it if necessary If accumulator is discharged the probe automatically turns off...

Page 9: ...web www tesla cz 9 Repairs Any repairs and non basic maintenance must be performed exclusively by TESLA manufacturer TESLA Podebradska 56 186 180 66 Prague 9 www tesla cz 10 Warranty This product is...

Page 10: ...er is included in package 12 Alternatives TSR3DNM probe is easy and quick to change in manufacturer to new type of these probes TSR3D WLAN Wireless and USB Radon Probe with Display TSR3DS SIGFOX Wirel...

Reviews: