TSR3D is designed for continuous measuring of radon concentrations in buildings and results are also 

shown  on display.

 

Portable probe basis is a measuring chamber with a semiconductor detector. Radon enters the chamber 

by diffusion through the input filter on the bottom of probe. The probe measures in autonomous and time 
continuous  way.  It  processes  results  every  4-minute  intervals  and  from  this  counts  short-term  moving 
average of radon concentration ( 1 hour moving average - average of 15 4-minute process intervals). It also 
counts long-term moving average of radon concentration ( 24 hours moving average). The probe saves time 
records  of  these  radon  concentration  values  including  values  of  humidity  and  temperature  within  its 
internal memory (typically at an interval of 1 hour). Next saved value is time record of measuring  energy 
spectrum  (typically  at  an  interval  of  12  hours).  All  results  are  shown  on  display  as  well  see  paragraph 
´Display Operation´. 

The probe is random for location in measured place, but generally it is put on the bottom of the probe. 

Bottom  of  the  probe  cannot  be  covered.  The  probe  can  be  switched  on/off  by  switch.  LEDs  „STAT“  a 
„CHRG“ indicate current status of probe see  ´Operation manual´ below.  

The  resulting  values  can  be  downloaded  continuously  during  measurement  or  at  once  at  the  end  of 

measurement.  Data  from  the  radon  probe  is  downloaded  to  a  PC  directly  via  USB  interface.  Capacity  of 
internal memory is almost unlimited. 

Summary of Contents for TSR3DNM

Page 1: ...ns Operation Manual v 1 2017 Table of Contents 1 Introduction 2 2 Description and Utilization 3 3 Scope of Delivery 4 4 Product Specification 5 5 Operating Instructions 6 6 Display Operation 6 7 Basic...

Page 2: ...asis of used product If you experience any problems with your product please contact us at TESLA Podebradska 56 186 180 66 Prague 9 www tesla cz 1 Introduction This document describes technical specif...

Page 3: ...The probe saves time records of these radon concentration values including values of humidity and temperature within its internal memory typically at an interval of 1 hour Next saved value is time re...

Page 4: ...ing results into internal memory The resulting values are downloaded after end of the measurement by B way B Probe connected via USB Using usbRADONview app and USB cable is possible to download result...

Page 5: ...ponse rate 30 minutes RaA 3 hours RaA RaC Radon records calculated from RaA quicker less sensitive calculated form RaA RaC slower more sensitive Measuring relative humidity range 10 90 Measuring tempe...

Page 6: ...accumulator is described in chapter Basic Maintenance Accumulator charging Power supply According to operation method the radon probe can be supplied 1 By internal accumulator for portable use Radon p...

Page 7: ...intro screen automatically after 5 seconds from intro screen directly from switched off display by pushing of button for longer than 1 second Changing of three data screens is cyclic without intro sc...

Page 8: ...cumulator charging During portable use of radon probe is essential to monitor state of internal accumulator and recharge it if necessary If accumulator is discharged the probe automatically turns off...

Page 9: ...web www tesla cz 9 Repairs Any repairs and non basic maintenance must be performed exclusively by TESLA manufacturer TESLA Podebradska 56 186 180 66 Prague 9 www tesla cz 10 Warranty This product is...

Page 10: ...er is included in package 12 Alternatives TSR3DNM probe is easy and quick to change in manufacturer to new type of these probes TSR3D WLAN Wireless and USB Radon Probe with Display TSR3DS SIGFOX Wirel...

Reviews: