background image

 

4/10 

TSR4M Radon Probe can be operated by these ways: 

 

A)

 

Standalone  probe

  -    Thanks  to  its  independent  battery  power,  portable  radon  measuring  probe 

supports flexible placing options within monitored structures. Accumulator will last for more than 
1  year  after  full  charging.  After  switching  on  TSR4M  immediately  starts  measuring  and  saving 
results into internal memory. The resulting values are downloaded after end of the measurement 
by B) way. 

 

B)

 

Probe  connected  via  USB

  –  Using  usbRADONview  app  and  USB  cable  is  possible  to  download 

results  to  PC  continuously  during  measurement  or  at  once  at  the  end  of  measurement. 
usbRADONview  application,  drivers  and  user  manual  is  free  downloaded  on  website: 

http://www.tesla.cz/en/ke-stazeni/

. 

 

 

 
 

3

 

Scope of Delivery 

 

TSR3DNM USB Radon Probe with display and big memory 

 

Power adapter 230VAC/5VDC 

 

USB cable A-B  

 

Probe holder 

 

Operation Manual 

 

 

Summary of Contents for TSR3DNM

Page 1: ...ns Operation Manual v 1 2017 Table of Contents 1 Introduction 2 2 Description and Utilization 3 3 Scope of Delivery 4 4 Product Specification 5 5 Operating Instructions 6 6 Display Operation 6 7 Basic...

Page 2: ...asis of used product If you experience any problems with your product please contact us at TESLA Podebradska 56 186 180 66 Prague 9 www tesla cz 1 Introduction This document describes technical specif...

Page 3: ...The probe saves time records of these radon concentration values including values of humidity and temperature within its internal memory typically at an interval of 1 hour Next saved value is time re...

Page 4: ...ing results into internal memory The resulting values are downloaded after end of the measurement by B way B Probe connected via USB Using usbRADONview app and USB cable is possible to download result...

Page 5: ...ponse rate 30 minutes RaA 3 hours RaA RaC Radon records calculated from RaA quicker less sensitive calculated form RaA RaC slower more sensitive Measuring relative humidity range 10 90 Measuring tempe...

Page 6: ...accumulator is described in chapter Basic Maintenance Accumulator charging Power supply According to operation method the radon probe can be supplied 1 By internal accumulator for portable use Radon p...

Page 7: ...intro screen automatically after 5 seconds from intro screen directly from switched off display by pushing of button for longer than 1 second Changing of three data screens is cyclic without intro sc...

Page 8: ...cumulator charging During portable use of radon probe is essential to monitor state of internal accumulator and recharge it if necessary If accumulator is discharged the probe automatically turns off...

Page 9: ...web www tesla cz 9 Repairs Any repairs and non basic maintenance must be performed exclusively by TESLA manufacturer TESLA Podebradska 56 186 180 66 Prague 9 www tesla cz 10 Warranty This product is...

Page 10: ...er is included in package 12 Alternatives TSR3DNM probe is easy and quick to change in manufacturer to new type of these probes TSR3D WLAN Wireless and USB Radon Probe with Display TSR3DS SIGFOX Wirel...