10
Calibration kit for EFT / Burst Capacitive coupling clamp INA 3425
– Connect CAS 3025 (or INA 265A – 50 Ω terminator / attenuator) and oscil
-
loscope to the termination bracket as shown in Fig 5
Fig. 5: CAS 3025 (INA 265A – 50 Ω terminator / attenuator) and oscilloscope
connected to termination bracket.
– the whole test setup might look like shown in Fig 6
Fig. 6: complete test setup for Burst / EFT coupling clamp calibration
– set Burst generator to 2 kV
– measure burst pulse waveshape and parameters at Coupling clamp output.
– turn the Burst coupling clamp in the other direction in order to verify / calibrate
the other path, and repeat the measurements.