background image

12

Calibration kit for EFT / Burst Capacitive coupling clamp INA 3425

3.2 Typical burst wave and parameters at Coupling clamp output

Summary of Contents for INA 3425

Page 1: ...1 CALIBRATION KIT FOR EFT BURST CAPACITIVE COUPLING CLAMP INA 3425 USER MANUAL 601 339A...

Page 2: ...Calibration kit for EFT Burst Capacitive coupling clamp INA 3425 CALIBRATION KIT FOR EFT BURST CAPACITIVE COUPLING CLAMP INA 3425 USER MANUAL...

Page 3: ...generator output 11 3 2 Typical burst wave and parameters at Coupling clamp output 12 4 Technical data 13 4 1 Technical parameters INA 3425 transducer plate 13 4 2 technical parameters INA 3425 termi...

Page 4: ......

Page 5: ...eeds to be followed during use of this accessory INA 3425 together with the EFT Burst generator and coupling clamp should only be installed and used by authorised and trained EMC specialists or calibr...

Page 6: ...e pulses are applied Up to pulse voltages up to ca 1 kV the user will not notice anything by getting in contact with the plates up to 4 kV he will notice some pain similar to what one feels during an...

Page 7: ...d for the calibra tion burst EFT coupling clamps as specified in basic standard IEC 61000 4 4 Ed 3 0 2012 The INA 3425 consists in a transducer plate and a termination bracket con nection adapter Fig...

Page 8: ...EN 61000 4 4 Ed 3 0 2012 2 3 Operating Set up Burst EFT generator oscilloscope and coupling clamp to calibrate on an adequate ground plane respect setup as specified in IEC 61000 4 4 Ed 3 0 2012 Use...

Page 9: ...Fig 3 CDN 3425 connected to the EFT output of a NSG 3040 generator Connect termination bracket connection adapter at the other end of the coupling clamp to calibrate as shown in Fig 4 Fig 4 connection...

Page 10: ...erminator attenuator and oscilloscope connected to termination bracket the whole test setup might look like shown in Fig 6 Fig 6 complete test setup for Burst EFT coupling clamp calibration set Burst...

Page 11: ...11 3 TYPICAL WAVEFORM DATA 3 1 Typical burst wave and parameters at generator output...

Page 12: ...12 Calibration kit for EFT Burst Capacitive coupling clamp INA 3425 3 2 Typical burst wave and parameters at Coupling clamp output...

Page 13: ...70 x 0 5 mm Total thikness 1 5 mm Dielectris strength isolation 5 kV Weight 0 3 kg approx 4 2 technical parameters INA 3425 termination bracket connecting adapter Input connector Banana Output connect...

Page 14: ...25 is a Calibration set for burts EFT It consists of INA 265A 50 1000 to 1 terminator attenuator and of INA 266 1000 2000 to 1 terminator attenu ator of a coax cable of 1 meter RG 58 and of a traceabl...

Page 15: ...mount of a mild non foaming household cleanser as well No chemicals should be used for cleaning purposes 5 2 Warranty Teseq grants a warranty of 24 months against material and manufacturing faults The...

Page 16: ...0 8460 8080 F 86 10 8460 8078 chinasales teseq com Germany Teseq GmbH T 49 30 5659 8835 F 49 30 5659 8834 desales teseq com Singapore Teseq Pte Ltd T 65 6846 2488 F 65 6841 4282 singaporesales teseq c...