EN
50
TROUBLESHOOTING
If the unit does not operate after taking these measures, contact the nearest dealer.
PROBLEM
POSSIBLE CAUSES
POSSIBLE SOLUTION
The unit cannot power on
Are the batteries exhausted?
Replace the batteries.
Are the batteries installed correctly?
Insert the batteries observing polarity.
Stimulation weak or cannot
feel any stimulation
Electrodes are dried out or dirty.
Replace with new electrodes.
Electrodes do not stick to skin well.
Replace with new electrodes.
Poorly positioned electrodes.
Switch off the device and reposition the electrode on your
skin then switch back on.
Stimulation is uncomfortable
Intensity is too high.
Decrease intensity.
Electrodes are too close together.
Reposition electrodes to be at least 2 inches apart.
Is the device being operated according to
the manual?
Please check the manual before use.