Bionova
®
MiniPro
Page 6
Safety information
Important
•Do not use this product in a manner not specified
by Terragene
®
S.A., otherwise the protection
provided by the product might be affected.
•Only Terragene
®
S.A. authorized personnel can
access or service the internal components of the
Auto-reader. Parts or components inside the Auto-
reader should not be manipulated by the user.
cords, and USB cable. Verify that all the included
elements are in good condition on daily basis. If any
of them is damaged, discontinue their use. The use
of cables, adapters, cords, and/or power supplies
different from the ones included may cause fires,
electrical shocks, or even physical injuries.
•Do not plug any devices into the Auto-reader’s
USB port other than a personal computer (PC).
The PC has to be compliant with IEC60950-1,
IEC 62368-1 or comparable, with safety extra-
low voltages on its USB ports. Ask a qualified
technician to verify device compatibility. Attaching
any other device to the USB port may damage the
Auto-reader and may not be safe for the user.
•Do not attempt to repair the Auto-reader by
yourself, that could lead to major and irreversible
damages to the device. In case of device
malfunction, contact your local distributor for
further assistance.
To reduce the risks associated with contaminated
surfaces with protein levels outside the user-
defined threshold.
•Please read, understand, and follow the Instructions
for Use of each Protein Pen before its incubation.
•Do not remove the Protein Pen before the Auto-
reader reports the final readout result. Check the
printed ticket.
To avoid a potentially hazardous situation:
•Avoid contact with the hot metal block inside
each incubation position.
•Do not insert your fingers, or any other element,
inside the incubation positions.
•Place only compatible indicators inside the
incubation positions.
Operating conditions
Power supply specifications
Input
parameters
Voltage range
Frequency
*Current
Output parameters
Voltage
**Current
Operating
conditions
(100-240)
50/60
0,5
Values
12
2
Units
AC Volts
Hertz
Amperes
Units
DC Volts
Amperes
* 1 Ampere in USA and Canada.
** 3 Amperes in USA and Canada.
Terragene
®
S.A. recommends the use of UPS instead of voltage
stabilizers, since they fulfill two functions: to stabilize and
maintain the energy during a power outage.
Environment operating conditions
Environmental
Conditions
Altitude
Operation temperature
Relative Humidity
Installation/overvoltage
Degree of contamination
Storage temperature
Voltage
Operating
Conditions
3500 (máx.)
10-30
30-80
category II
2
10-30
12
Units
Meters
Celsius
%
Celsius
DC Volts
Bionova
®
MiniPro Auto-reader complies with the following
standards and directives:
Device Safety Compliance
Electrical Safety.
Electromagnetic
Compatibility (EMC).
European Commission.
IEC 61010-1
IEC 61010-2-010
Low Voltage Directive
2014/35/EU.
EN 61326-1
EMC Directive 2014/30/EU
RoHS Directive 2011/65/EU.
WEEE Directive 2012/19/EU.