3.4
Refrigeration Module
The optional Refrigeration Module quickly and quietly cools
the internal chamber to the user’s specifications. The
module is mounted on top of the “dirty” side of the pass-
through and requires that a support frame be installed
below the pass-through (see
Figure 4
).
Be sure the evaporator drain tube exits from the bottom of
the pass-through (see
Figure 21
). Connect the drain tube
to an appropriate drainage device.
Turn on the unit using the green master power switch on
the “dirty” side of the pass-through. The temperature
controller is preconfigured for basic operation prior to
shipment.
To prevent overheating of the compressor and evaporator coils, the refrigeration module’s air intake and
exhaust vents must be clear of obstruction after installation.
CAUTION
To enter a new set-point:
1.
Press MENU once to open Setpoint 1 (Press MENU twice for Setpoint 2)
2.
Display flashes the first digit of Setpoint 1
3.
Press MAX to set the digit, 0-9
4.
Press MIN to activate the next digit (flashing)
5.
Repeat steps 3 and 4 until all digits are set
6.
Press ENTER to save the new setting and return to Run Mode
For additional instructions on setting alarms or reconfiguring the controller, refer to the manual from the controller manufacturer:
http://www.omega.com/Manuals/manualpdf/M2491.pdf
Specifications
Condenser
Evaporator
Refrigerant
R-22
Btu/hr @ 10° TD
1,200
Horsepower
¼ HP
Btu/hr @ 12° TD
1,440
Btu/hr
3,660
Btu/hr @ 15° TD
1,800
Power Requirements
115V, 60Hz, 1Ph
CFM
210
Defrost
Electric
Drain Connection
½” O.D.
Fan Diameter
5-1/2”
Number of Fans
2
Volts
115
Figure 21:
Drain tube exiting the evaporator inside the pass-through