Tera TOM 584 Manual Download Page 7

 

 

7

5.3 

Start the Software 

 

Connect the 

TERA

 – Ohmmeter with the PC – Interface cable to a free serial PC Interface  

(COM 1 or COM 2) from your PC, and press key „A“ to switch on the TERA – Ohmmeter. 
Start under Windows your EXCEL Program.  
Open in Excel the file „UT584.xls“ in the directory  c:\tom584 

 

For more information about this program use the HELP bottom 

 
 
 

5.4 Software 

Manual 

 

To select the Menüpunkte use Coursor - Buttons 

↑↓

. With the  <Enter> Button Menüpunkt will be started. 

 

Change start condition 

 

 

Timer ?  

(0-240s) 

give in direct the time for the count-down (0 – 240 sec). 

 

Average ? 

 [y/n]   

with “y” the mid-value (Average) of all the measurements, during 

 

  

 

 

 

the count-down timer is running, are calculated. 

 

  

 

 

 

with „n” the last measured value is shown in the display (Lapse). 

 

 

Important

 !

   The changed items are not stored in the EEPROM ! 

 

Start Measurement   

 

 With

 <Enter> the count-down timer for the measurement cycle will be started. After the count-down 

timer returned to zero it is possible to store the complete measurement-curve. 
(100ms – sample-time). 

 

 

Save Data ?  

[y/n]   

with “y” the measured values will be stored in the file MV.DAT .  

 

  

 

 

 

These File (MV.DAT) can be imported and modified with a 

  

    spread-sheet 

analysis 

(ex. 

EXCEL). 

 

  

 

 

 

For separation of the data-cells a comma is used. 

 

 

 

Important

 !

 

With each saving the file MV.DAT will be overwritten ! 
If you want to store different files, you need to rename the actual file MV.DAT. 
Then you can store the next MV.DAT. 

 

Save Data - Files from TOM 

 

 with

 <Enter> the complete data-memory (in the EEPROM of the TOM) are transmitted and stored 

 

as the file “FILE.DAT”. The modification- procedure is like for the “MV.DAT”. 

 

Reset Tera 

 

 with

 <Enter> a reset the processor in the TOM will be started and the selected values in the memory 

 

of the TOM are transmitted to the PC. 

 

Exit 

 

with <Enter> the programm will be closed. 

Summary of Contents for TOM 584

Page 1: ...TF The TERA Ohmmeter TOM 584 works with the Current Voltage Measurement This is the inexpensive version of the TOM 374 High resistances up to 2x1012 Ohm 2 Tera Ohm can be measured with 100 volts and...

Page 2: ...Set Up Functions 5 3 1 Timer 6 4 Other Display Information s 6 4 1 Display Contrast 6 5 PC Operating 6 5 1 System requirements 6 5 2 Softwareinstallation 6 5 3 Starting the Software 7 5 4 Software Ma...

Page 3: ...ith 1200 V and then measured the discharge time in an ionization air The start and stop voltage can be selected The pos and neg discharge time as well as the offset voltage will be shown in the displa...

Page 4: ...vered with a 9V NiMH Battery and is directly ready to start Only the enclosed plug power supply should be used for charging the battery and stationary work The electrodes are connected to the correspo...

Page 5: ...me and in the display appears AVERAGE LAPSE By pressing key A the middle value over the measure time will be displayed AVERAGE By pressing key B the last value after the end of the measure time will b...

Page 6: ...t can quietly be completed If the Battery Voltage is lower as 7 0 volts following display appear one after the other Then the unit automatically switches off to avoid a destroying of low Voltage of th...

Page 7: ...EPROM Start Measurement With Enter the count down timer for the measurement cycle will be started After the count down timer returned to zero it is possible to store the complete measurement curve 100...

Page 8: ...ght app 350 g Display alphanumeric Display 1 line with 16 Letters Display field 60 mm x 15 mm Measure Ranges Resistor 20 k 2 0 T Tolerance 1xEXX PC Interface serial Interface 9 pin SUB D Connector PC...

Page 9: ...EOS ESD S 11 11 DIN EN 61340 5 1 2 3 comprising to the measurement of transit and surface resistances 1 shielded and 1 no shielded Measure cord Other electrodes are available on inquiry Info Make sure...

Page 10: ...10 12 Legend 1 Power Supply connector 9 12V DC 2 Serial PC Interface 3 Measure Voltage Output 4 Measure Input 5 Contrast Potentiometer for the LC Display 6 Ground connector...

Page 11: ...ezieht die Forderung der EWG Richtlinie ber die elektromagnetische Vertr glichkeit Richtlinie des Rates vom 3 5 1989 89 336 Stand Juli 1993 erf llt insbesondere der Normen EN 61010 Safety EN 55011 Gru...

Reviews: