3.1 Timer

 

 

If the timer was not activated, the TOM 584 directly switches on the resistance display in the on-line 
operation. In the display appears: 

 

 

  

 

 

 

 

=5.5xE10

    T=OFF 

 

 

 
 

4. 

Other Display Information’s 

 

If the maximum measurement is exceeded appears in the display:  

 

     >2.0xE12 

(T=OFF) 

 

 

If the minimal measurement becomes under-stride appears in the display:  

 

     <2.0xE04

   T=OFF  

 

If the Battery Voltage is lower than 7,6 Volt, report the unit in the change with the current  line:  

 

 

  

 

 

 

  

LOW BATTERY ! 

 

The accumulator must be charged again, however the current measurement can quietly be completed. If the 
Battery Voltage is lower as 7,0 volts, following display appear one after the other. Then the unit 
automatically switches off, to avoid a destroying of low Voltage of the battery. 

 

 

  

 

 

 

  

LOW BATTERY  

 

  

 

 

 

      AUTO OFF 

 

  

 

 

              SWITCHING OFF 

 
 

4.1 

Display contrast  

 

On the right side of height of the display there is a hole trough it you can screw a potentiometer for adjust 
Contrast of the LCD - Display. 

 
 
 
 
 

5. 

PC - Operating 

 

5.1 

System requirement  

 

• 

PC with 486er CPU or higher 

• 

3,5” Floppy disk drive 1,44MB (HD) 

• 

free serial Interface COM1 or COM2 

• 

Windows98, 2000, XP with EXCEL 

 

 

5.2  

Installations  

 

Put the disk with the TOM software into the floppy drive. Go to „Start“ then Execute. Type Install and press 
<Enter>. Now the software will be installed.

 

 

Summary of Contents for TOM 584

Page 1: ...TF The TERA Ohmmeter TOM 584 works with the Current Voltage Measurement This is the inexpensive version of the TOM 374 High resistances up to 2x1012 Ohm 2 Tera Ohm can be measured with 100 volts and...

Page 2: ...Set Up Functions 5 3 1 Timer 6 4 Other Display Information s 6 4 1 Display Contrast 6 5 PC Operating 6 5 1 System requirements 6 5 2 Softwareinstallation 6 5 3 Starting the Software 7 5 4 Software Ma...

Page 3: ...ith 1200 V and then measured the discharge time in an ionization air The start and stop voltage can be selected The pos and neg discharge time as well as the offset voltage will be shown in the displa...

Page 4: ...vered with a 9V NiMH Battery and is directly ready to start Only the enclosed plug power supply should be used for charging the battery and stationary work The electrodes are connected to the correspo...

Page 5: ...me and in the display appears AVERAGE LAPSE By pressing key A the middle value over the measure time will be displayed AVERAGE By pressing key B the last value after the end of the measure time will b...

Page 6: ...t can quietly be completed If the Battery Voltage is lower as 7 0 volts following display appear one after the other Then the unit automatically switches off to avoid a destroying of low Voltage of th...

Page 7: ...EPROM Start Measurement With Enter the count down timer for the measurement cycle will be started After the count down timer returned to zero it is possible to store the complete measurement curve 100...

Page 8: ...ght app 350 g Display alphanumeric Display 1 line with 16 Letters Display field 60 mm x 15 mm Measure Ranges Resistor 20 k 2 0 T Tolerance 1xEXX PC Interface serial Interface 9 pin SUB D Connector PC...

Page 9: ...EOS ESD S 11 11 DIN EN 61340 5 1 2 3 comprising to the measurement of transit and surface resistances 1 shielded and 1 no shielded Measure cord Other electrodes are available on inquiry Info Make sure...

Page 10: ...10 12 Legend 1 Power Supply connector 9 12V DC 2 Serial PC Interface 3 Measure Voltage Output 4 Measure Input 5 Contrast Potentiometer for the LC Display 6 Ground connector...

Page 11: ...ezieht die Forderung der EWG Richtlinie ber die elektromagnetische Vertr glichkeit Richtlinie des Rates vom 3 5 1989 89 336 Stand Juli 1993 erf llt insbesondere der Normen EN 61010 Safety EN 55011 Gru...

Reviews: