WX-1 Series Instrument Revised: 10.3.19
pg. 14
Follow these instructions to update the calibration for an existing web on
the menu or to add a calibration for a new web. If you would like to add a
new web, first create a name for it by following the instructions under
“Custom Names” in Section 6, above.
1. Scroll to
SETUP
and press ENTER. Next, scroll to
RE-CAL TENSION
and press
ENTER. The screen will now display:
SETUP, RE-CALIBRATE TENSION,
ENTER PASSWORD, CAL DATA WILL BE LOST!
2.
Enter Password
. Contact Tensitron at
3. The display now indicates:
SETUP RE-CALIBRATE TENSION
with a list of the
calibration names and custom names programmed into the instrument. Scroll to
the description you wish to recalibrate, highlight it and then press
ENTER
.
4. The next screen will indicate:
CALIBRATE TENSION, NUMBER OF POINTS.
Using the up and down arrows select the number of calibration points to which you
will be tensioning the web. The minimum is 5 points. Zero is a calibration point
already set, so you will need to define a minimum of 4 increasing tension values
such as: 50, 100, 150 and 200. If desired you can define more than 5 weight
points, with a maximum of 10 points total. If 10 calibration points were selected the
entries could look something like this: 10, 20, 30, 40, 50, 60, 70, 80, 90.
Figure 7: Free hanging weight
simulating web tension.