TensioMed® Arteriograph24™
17
2.4.3. Error codes for users
The error codes, which appear on the display and their meanings, are described below.
Please, note that you should not make any conclusions if an error appears once because
the movements of the patient can imitate several types of errors. If the device cannot
measure the blood pressure (e.g. because of movement), the measurement will be
interrupted. With the TensioMed® TensioWin™ program - in case of a faulty measurement
- it is possible to set the device to repeat it according to the measuring plan after 1 minute.
The meanings of the error codes shown by the device are as follows:
1
The device could not measure the patient’s blood pressure within the
measuring time
3
The measurement was interrupted due to the weakness of the battery
31
The cuff is not connected to the device
32
The cuff tube is broken or something got into the tube (e.g. water).
33
The cuff (or device) is leaking
34
The cuff is not on the patient‘s arm
35
The measurement was interrupted for some reason (e.g. because the patient
pressed the button).
37
During the blood pressure measurement the cuff pressure has reached or
exceeded the maximum allowable pressure of 300mmHg 's value.
91‐97
The BP measurement was not successful due to the failure of the device or the
batteries are weak
100
The measured result cannot be considered as a real BP value or the patient has
arrhythmia
101
The measuring circumstances e.g. the moving of the patient, disturbs the
measurement
102
The device cannot sense the HR
110
The measured result cannot be considered as a real BP value because of some
movement or arrhythmia
111
systole > max. inflation
115
The HR cannot be calculated or cannot be considered as a real HR value
116
Not enough evaluable results either from sys or from dia.