14
Technical specifications for Transcutaneous Electrical Nerve Stimulator (TENS)
mode
Technical specifications for Electrical Muscle Stimulator (EMS) mode
Waveform
Mono-phase square pulse wave
Pulse amplitude
Adjustable, 0-105mA peak at 1000 ohm Load each channel.
1 mA/Step.
Pulse Width
Adjustable, from 50 to 300µS, 10µS/step
Pulse Rate
Adjustable,, from 1 to 150Hz, 1 Hz/step
contraction time Adjustable,1-30 seconds , 1 Sec. / step
Relaxation
(OFF) time
Adjustable. 0
—60 seconds , 1 Sec./ step
Ramp time
Adjustable. 1-6 seconds, 1 Sec
. / step, The “On” time will increase
and decrease with the setting value.
Synchronous (S)
Stimulation of both channels occurs synchronously. Th
e “ON’ time
includes
“Contraction”, ‘Ramp Up” and “Ramp Down” time,
ON TIME =Cont
Ramp up+ Ramp down
Alternate (A)
The Stimulation of CH2 will occur after the 1st working of CH1 is
completed. OFF time should = or > ON time
ON TlME=Contr Ramp up + Ramp down.
OFF TIME>ON TIME
Delay {D)
The Stimulation of the CH2 will occur after the 1st working of CH1 is
s Delay Time. Delay time is adjustable from 1 to 10 Sec.
ON TlME=Contr Ramp up + Ramp down
Waveform
Mono-phase square pulse wave
Pulse amplitude
Adjustable, 0-105mA peak at 1000 ohm Load each channel.
1 mA/Step.
Pulse Width
Adjustable, from 50 to 300uS, 10uS/step
Pulse Rate
Adjustable,, from 1 to 150Hz, 1 Hz/step
Burst (B)
Burst rate: Adjustable, 0.5-5Hz, 0.1 Hz/step
Pulse width adjustable, 50-300µS
Frequency fixed = 100 Hz
Normal (N]
The pulse rate and pulse width are adjustable.
Generates continuous stimulation based on the setting value.
Pulse Width
Modulation (M)
The pulse width is automatically adjusted in a cycle time
The pulse width is decreased from its original setting to 60% in its
set cycle time, and then increased from 60% to its original setting in
next set cycle time. In this program, pulse rate (1 to 150Hz), pulse
width (50 to 300µS) and cycle time (5 to 30sec) are fully adjustable.
Pulse Rate
Modulation(M1)
The pulse rate is automatically varied in a cycle time. The pulse rate
is decreased from its original setting to 60% in setcycle time, and
then increased from 60% to its original setting in next cycle time. In
this program, pulse rate (1 to 150Hz), pulse width (50 to 300µS)
and cycle time (5 to 30sec) are fully adjustable.