TM-6001
9
6 Electrical specifications:
To ensure accuracy the ambient temperature should
be 23°C ± 5°C with a humidity of 80% RH (maximum)
non-condensing. In addition, perform a Zero
adjustment after each range change.
±(0.8
%
reading
+
10digits)
Resistance measurements
Temperature coefficient
:
±(0.1
%
rdg
+
0.5digits)/
℃
.
Measurement current frequency
:
1KHZ±30HZ.
Measurement open-circuit terminal voltage
:
7Vp-p
CAUTION
The maximum input for DC voltage is
60V (No AC voltage Input permitted). Do
not attempt to measure high voltages to
avoid electrical shocks or damages to the
instrument.
。
Range
Resolution
Measurement
current
Accuracy
4mΩ
1μΩ
150mA approx.
±(1
%
reading
+
20digits)
40m
Ω
10μΩ
150mA approx.
±(0.8
%
reading
+
10digits)
400m
Ω
100μΩ
50mA approx.
4
Ω
1mΩ
15mA approx.
40
Ω
10mΩ
1.5
m
Aapprox.
400
Ω
100mΩ
150
μ
A approx.