TM-6001
14
7. Press Zero adjustment
key for 2 seconds
to start the zero adjustment. A flashing “0 ADJ”
appears on the screen; when the tester reads a
steady value of the resistance lower than 1000;
then the screen displays “0” and stop flashing ”0
ADJ”.
Press
key again to disable the zero
adjustment.
8. Connect the red test probe to the positive battery
terminal, and the black test probe to the negative
battery terminal.
9. Read the battery internal impedance or DC voltage
directly and Temperature measurement from the
display.
Note
:
When the measured DC voltage or battery
internal impedance value is over range, “OL”is
displayed.