RESET
If pressed prior to a scan, will return measurement cursors to their
position during the previous scan. Used for other miscellaneous
operations, such as setting the time clock.
Stage Keys
These include the 4 arrow keys and the NXT (next) key in the center.
The arrows indicate the direction in which the stage, (not the stylus),
will move when pressed. They are used to move the sample stage and
during stage programming. The stage has two speeds in each direction:
0.13 mm/sec and 13 mm/sec. Pressing the NXT key before the desired
arrow key causes the stage to move at the faster rate.
RANGE
Allows the user to set the scan length/sample density. The
highlighted values change as either the ^ or “down arrow” key is
pressed.
PLOT
Used to zoom in on a section of a profile between the measurement
cursors, or to level a profile.
CUR
This key and its associated <- and -> keys allow the operator to set scan
direction and place the measurement cursors.
LEVEL
Used to level the baseline of a profile.
TABLE
The ^ and “down arrow” keys will move the sample chuck up and down
respectively.
Guess.
FEED
Printer paper feed.
START/STOP
Starts/stops the scan.
3. Operating Procedure
3.1. Setting Scan Parameters
Up to 9 scan menus containing scan parameters can be stored in the system. It is
usually convenient, however, to simply bring up the default menu and modify it to suit
your needs, as follows.
1. Make sure the power is on. Turn up the brightness.
2. Press VID to bring up a menu.
3.
Press ENT to initiate "Enter Mode". You will see a message at the bottom of the