INSTALLATION
telemark.com
23 of 48
Rev 1.0.0
3.4.6 Host
–
J1
The host port is for upgrading the sweep software.
IF J1 IS CONNECTED TO A PC
THE DIGITAL SWEEP WILL AUTOMATICLY GO IN TO DOWNLOAD MODE.
Figure 3-7, Host Port
3.4.7 Output - J3
Connection of the sweep generator to the EB Source is shown in Figure 3-D, 3-E, 3-
F, 3-G or 3-H. The horizontal and lateral coils should be brought out of the tank by
way of a feedthrough and connected to pins 1, 2, 3 of J3 on the Sweep Module as
shown. The interconnecting wire must be capable of passing a minimum of two
amperes. The return wire is shared by both longitudinal and lateral coils. The sweep
voltage is grounded inside the sweeper. However, you should connect the return wire
to ground at the EB source end. To leave the return wire ungrounded could damage
the sweeper. Normally one side of each coil is connected to ground at the electron
beam source. The return wire is connected to the same spot inside the tank.
Figure 3-8, Telemark standard Coil Wire Reference Code