3 Troubleshooting
Teledyne Analytical Instruments
3-2
Symptom
Symptom
Symptom
Symptom
Symptom
The oxygen reading fluctuates
or appears to be incorrect.
No display.
Unit reads 00.
What to Do
What to Do
What to Do
What to Do
What to Do
A) During calibration, make
sure there are no restric-
tions on exhaust side of
sensor. If the reading
changes with flow, the sen-
sor is pressurized or there
may be a leak in the sys-
tem.
B) If a high degree of accuracy
is desired, or the concentra-
tion of O
2
is in excess of
40%, calibration with 100%
is recommended.
C) Do not hold sensor or sub-
ject it to changes in tem-
perature during calibration.
D) If humidified gas is used to
ventilate the patient, water
vapor actually dilutes the
gas. See Chapter 2.0: Hu-
midity, Temperature.
E) If a blender is used, check
its calibration. See Chapter
2.0: Discrepancy in Read-
ings.
A) Check/replace batteries.
B) Check battery connections.
A) Check and reconnect, then
calibrate.
Why
Why
Why
Why
Why
Like all O
2
sensors, the T-7
detects the changes in the
partial pressure of O
2
. See
Operations: Pressure.
Quick changes in temperature
can produce temperature
tracking errors.
A) Batteries expired.
B) Bad battery connection.
A) Sensor not plugged in.
NOTE:
NOTE:
NOTE:
NOTE:
NOTE: In the event that none of these procedures produce desired results, remove the
In the event that none of these procedures produce desired results, remove the
In the event that none of these procedures produce desired results, remove the
In the event that none of these procedures produce desired results, remove the
In the event that none of these procedures produce desired results, remove the
batteries and return the unit to Teledyne for repair.
batteries and return the unit to Teledyne for repair.
batteries and return the unit to Teledyne for repair.
batteries and return the unit to Teledyne for repair.
batteries and return the unit to Teledyne for repair.
NOTE:The TED 60T Service Manual (P/N SNC40575B.03) is available from the factory and
NOTE:The TED 60T Service Manual (P/N SNC40575B.03) is available from the factory and
NOTE:The TED 60T Service Manual (P/N SNC40575B.03) is available from the factory and
NOTE:The TED 60T Service Manual (P/N SNC40575B.03) is available from the factory and
NOTE:The TED 60T Service Manual (P/N SNC40575B.03) is available from the factory and
contains circuit diagrams, test procedures and other information which assist
contains circuit diagrams, test procedures and other information which assist
contains circuit diagrams, test procedures and other information which assist
contains circuit diagrams, test procedures and other information which assist
contains circuit diagrams, test procedures and other information which assist
qualified technical personnel in repairing the Analyzer.
qualified technical personnel in repairing the Analyzer.
qualified technical personnel in repairing the Analyzer.
qualified technical personnel in repairing the Analyzer.
qualified technical personnel in repairing the Analyzer.