Over Voltage Protection Setup
Background
Over Voltage Protection (OVP) protects T3PS36006
and DUT from excessive output Voltage. The user
sets the maximum output voltage limit before
operation. When the output voltage exceeds this
limit, the indicator shows the over voltage status
and the output is shut off immediately.
OVP setup
1.
Slide the rear panel switch to
the “O.V.P. SET” position.
C3
C
C1
NORAL
O.V.P. Set
2.
The OVP indicator on the front
panel turns green, indicating
OVP setup.
O.V.P.
SET
TRIP
3.
The Voltage meters show the OVP setting level
instead of the output level. The Current meters
show zero (0.00).
CH1 / CH3
CH1 / CH3
4.
Adjust the OVP level using the
rear panel terminal. The
setting on the front panel
meter changes accordingly.
C1
INCREASE
DECREASE
Setting range
Channel 1
1.0V ~ 67.0V
Channel 2
1.0V ~ 67.0V
Channel 3
0.1V ~ 6.0V