Operating Instructions
Teledyne ML – T101 Operation Manual
64
4.4.
SETUP MODE
The
SETUP
mode allows you to configure the analyzer’s hardware and software features,
perform diagnostic procedures, gather information on the instrument’s performance and
configure or access data from the internal data acquisition system (DAS). For a visual
representation of the software menu trees, refer to Appendix A.
Pressing the
SETUP
button activates a prompt for a security password. The default
password is
818
. Press ENTR to proceed.
However, there is the option to enable a higher level of security; described in Section
4.4.5.
Other password levels exist, allowing access to special diagnostic tools and variables used
only for specific and rarely needed troubleshooting and adjustment procedures. They
may be made available as needed by Teledyne ML’s Technical Support department.
The following two tables decribe the menus under Setup mode:
Table 4-4. Primary Setup Mode Features and Functions
MODE OR FEATURE
TOUCHSCREEN
BUTTON
DESCRIPTION
MANUAL
SECTION
Analyzer Configuration
CFG
Lists key hardware and software configuration
information
Auto Cal Feature
ACAL
Used to set up and operate the AutoCal feature.
Only appears if the analyzer has one of the internal
valve options installed
Internal Data Acquisition
system (DAS )
DAS
Used to set up the DAS system and view recorded
data
Analog Output Reporting
Range Configuration
RNGE
Used to configure the output signals generated by
the instrument’s Analog outputs.
Calibration Password
Security
PASS
Turns the calibration password feature ON/OFF
Internal Clock
Configuration
CLK
Used to Set or adjust the instrument’s internal
clock
Advanced SETUP
features
MORE
This button accesses the instrument’s secondary
setup menu
Table 4-5. Secondary Setup Mode Features and Functions
MODE OR FEATURE
TOUCHSCREEN
BUTTON
DESCRIPTION
MANUAL
SECTION
External Communication
Channel Configuration
COMM
Used to set up and operate the analyzer’s various
external I/O channels including RS-232; RS 485,
modem communication and/or Ethernet access.
System Status Variables
VARS
Used to view various variables related to the
instrument’s current operational status
System Diagnostic
Features
DIAG
Used to access a variety of functions that are used
to configure, test or diagnose problems with a
variety of the analyzer’s basic systems
Summary of Contents for T101
Page 2: ......
Page 3: ......
Page 4: ......
Page 6: ...ii This page intentionally left blank ...
Page 10: ...vi This page intentionally left blank ...
Page 12: ...viii This page intentionally left blank ...
Page 22: ...TABLE OF CONTENTS Teledyne ML T101 Operation Manual xviii This page intentionally left blank ...
Page 166: ...Remote Operation Teledyne ML T101 Operation Manual 162 This page intentionally left blank ...
Page 248: ......
Page 336: ...This page intentionally left blank B 4 07266B DCN6485 ...
Page 339: ...APPENDIX D Wire List and Electronic Schematics D 1 07266B DCN6485 ...
Page 340: ...This page intentionally left blank 07266B DCN6485 D 2 ...
Page 345: ...D 7 07266B DCN6485 ...