089770000 Rev 05
Teledyne API N500 CAPS NO
X
Analyzer with NumaView™ Software
85
S
AMPLE
C
ONDITIONER
R
EPLACEMENT
The conditioner assembly includes a dryer, an AREF filter, and a charcoal filter.
Change the dryer component as follows:
1. Turn off power to the analyzer and pump and remove the power cord and the analyzer
cover.
2.
Gently
disconnect all tubing that extends out of the dryer assembly.
Note
Do not twist any of the white plastic fittings on the dryer:
they connect the inner drying tube to the outer purge tube
and are delicate. See Section 5.6.2.
3. Make note of the dryer’s orientation on the bracket in order to correctly position the
replacement dryer.
4. Cut the tie wraps that hold the dryer to the mounting bracket and take out the old dryer.
•
If necessary, unscrew the two mounting screws on the bracket and take out
the entire assembly.
5. Attach the replacement dryer to the mounting bracket in the same orientation as the old
dryer.
6. Fix the dryer to the bracket using new tie wraps.
7. Cut off excess length of the wraps.
8. Put the assembly back into the chassis and tighten the mounting screws.
9. Use at least two wrenches to carefully re-attach the tubes.
Note
Avoid twisting the dryer’s white plastic fittings, as this will
result in large leaks, which are difficult to trouble-shoot or fix.
10. Carry out a detailed leak check (see Section 5.6.5.1),
11. Close the analyzer and reconnect the power cord.
12. Power up pump and analyzer and re-calibrate the instrument after it stabilizes.