SETUP>DIGITAL OUTPUTS (OPTION)
Specify the function of each digital output (connected through the rear panel Digital I/O
connector, Figure 2-5) by mapping the outputs to a selection of “Signals” present in the
instrument. Create custom “Signals” in the Setup>Events menu (Section 2.5.2). The three
Relays can also be connected, and functions assigned.
To map Digital Outputs to Signals, select a pin in the Outputs list (DO1 thru DO5), then
make a selection from the Signals list and press the Map button; if/as needed, change the
polarity by pressing the Polarity button. Save any changes by pressing the Apply button,
or discard the changes by pressing the Home or the back button (a pop-up provides a
warning that the changes will be lost, and will prompt for confirmation to apply changes
or not). Map the Digital Relays in the same manner.
Navigate to the Utilities>Diagnostics>Digital Outputs menu to change the state
(ON/OFF) of individual digital outputs.
Figure 2-32. Digital Outputs Setup