View/edit the Communications parameters in the Setup>Comm>COM1 menu.
RS-232
•
Baud rate
: 115200 bits per second (baud)
•
Data Bits
: 8 data bits with 1 stop bit
•
Parity
: None
RS-485
•
Baud rate
:19200 bits per second (baud)
•
Data Bits
: 8 data bits with 1 stop bit
•
Parity
: None
PNEUMATIC CONNECTIONS
This section provides pneumatic connection and setup instructions for basic and valve
option configurations. Pneumatic flow diagrams are shown in Section 2.3.3. Calibration
instructions are provided in Section 4.
Before making the pneumatic connections, carefully note the following cautionary and
special messages:
CAUTION – General Safety Hazard
Do not vent calibration gas, exhaust gas or sample gas into
enclosed areas.
CAUTION – General Safety Hazard
In units with a permeation tube option installed, vacuum pump
must be connected and powered on to maintain constant gas
flow though the analyzer at all times. Insufficient gas flow
allows gas to build up to levels that will contaminate the
instrument or present a safety hazard to personnel.
Remove the permeation tube when taking the analyzer out of
operation and store in sealed container (use the original
shipping packaging).
(See Section 5.6.4 for instructions on how to remove the
permeation tube when the unit is not in operation).