NDIR Gas Analyzer
Maintenance 6
Teledyne Analytical Instruments., - Rev. 3
67
SETUP: CAL DELAY
Screen from the SETUP menu. It indicates the time
specified for delaying the onset of a calibration. This
delay is generally used for purging the cell before
calibrating.
Delay time can be changed using the front panel UP and
DOWN keys.
SETUP: = CHART X
Screen from the SETUP menu. It displays the current
value of full scale for the indicated range (x = 1–4).
Use the UP and DOWN keys to change the full scale
value for that output range.
SETUP: = ALARM L
Screen from the SETUP menu. It displays the current
setpoint for the LOW alarm.
Use the UP and DOWN keys to change the setpoint.
SETUP: = ALARM H
Screen from the SETUP menu. It displays the current
setpoint for the HIGH alarm.
Use the UP and DOWN keys to change the setpoint.
SETUP: 02 CAL GAS
Screen from the SETUP menu. It indicates the entered
value for the concentration of the oxygen span gas.
Value can be changed using the UP and DOWN keys in
the front panel.
MODE: CALIBRATE?
Screen from the MODE menu. It allows the user to
initiate a manual calibration – zero, span or both.
MODE: CHART = X
Screen from the MODE menu. Allows the user to select
an output range or enable the autorange feature.
MODE: FILTER = X
Screen from the MODE menu. Used to select the
amount of filtering applied to the measurement. See
Section 2.3.3.
MODE: CAL =
HOLD/TRACK
Screen from the MODE menu. Allows the user to select
whether the instrument TRACKs the concentration
during a calibration or HOLDs the last value before
calibration began.
MODE:
LINEAR/NONLINEAR
Screen from the MODE menu. Selects whether data is
linearized, nonlinearized, or displayed as a raw signal
from the detector.
MODE: BRIGHT = X
Screen from the MODE menu. Alters the display
brightness.
MODE: AUTO =
ZERO/FULL
Screen from the MODE menu. Used to determine the
type of auto calibration.