NDIR Gas Analyzer
Operation 4
Teledyne Analytical Instruments., - Rev. 3
41
•
Enter the desired range or select autorange.
Press
the MODE button repeatedly until you reach the
CALIBRATE? submenu. Press and
hold
the MODE
button one more time to lock onto the CHART
submenu.
There are 2 mutually exclusive choices to make from
this submenu: FIXED range or AUTO. If you are using
a FIXED range, the top line of the display will indicate
the currently selected range (1–4). See the example
screen on the left. When FIXED is in effect, you can
change to the next range by pressing the UP button or
enter the AUTO mode by pressing the DOWN button.
If AUTO range is selected, the display will indicate
AUTO. You can change back to FIXED range by
pressing the DOWN button. Pressing the UP button will
also place you in FIXED mode and will select the range
indicated on the lower left in the display.
3.
F
ILTER
The FILTER submenu allows the operator to choose the
amount of filtering that is applied to the measurement. There
are 3 filter settings available: 1, 10, 100 (10, 100, 1000 for
trace instruments). The response time of the analyzer is related
to the filter number. Choosing a lower filter number results in a
faster response, however higher filtering yields a smoother
measurement.
•
Enter a filter value.
Press the MODE button
repeatedly until you reach the CHART submenu. Press
and
hold
the MODE button one more time to lock onto
the FILTER submenu. Use the UP or DOWN arrows to
toggle through the available filter settings. The filter
value will be indicated on the top line of the display.
Releasing the MODE button will set the filter value.