NDIR Gas Analyzer
Operation 4
Teledyne Analytical Instruments., - Rev. 3
35
3.
E
NTER
A
UTOMATIC
C
ALIBRATION
F
REQUENCY (If enabled)
Press and hold the SETUP button to enter the number of hours
between automatic calibrations. Automatic calibrations
requires that the analyzer be connected to externally powered
solenoid valves. The type of automatic calibration (either zero
or full) is selected in the MODE menu.
NOTE: In a dual analyzer this menu selection will not appear
in the second IR channel (referred to as “slave”) and is also
not present in the battery portable model.
4.
D
EFINE
R
ANGES
Press and hold the SETUP button to enter the Range 1
submenu. If necessary, cycle through the submenus by
repeatedly pressing the SETUP button until the CALGAS
menu appears. Then press and
hold
the SETUP button once
more to enter the CHART 1 submenu.
NOTE:
The dynamic range of the IR7000 detector is 1000:1 which means
that it can respond to a 1 PPM change in a sample containing 1000
PPM. This is fixed by the design of the detector and is independent of
the range setting. The 4 user definable ranges are implemented
mainly to accommodate standard output devices such as chart
recorders. The full output 0–1V, 0–5V or 0–10V (or optionally, 4–
20mA) will be linearly scaled over the defined chart range.
•
Enter Range 1.
Use the UP and DOWN arrows on the
front panel while holding in the SETUP button to enter
or change the upper bound of the CHART 1 range.
NOTE:
The 4 user definable ranges must be consecutively increasing. That
is, the upper bound on each successive range must be greater than
the previous range’s upper bound. Range 4
≥
Range 3
≥
Range 2
≥
Range 1.
•
Save changes.
As before, once you release the SETUP
button you will be prompted to save or else lose any
changes you made. There is a 60-second timer on this
screen. Press the UP button to save your changes or the
DOWN button to discard the changes within the 60-
second time limit. A confirmation screen will then
appear.
•
Set the other ranges.
If desired, toggle to the next
submenu (CHART 2) and continue in a similar fashion