page 31
Device Status
When requested to transmit its status the Model 2002 responds with a five digit number which is
explained in the following:
digit#: 1
2
3
4
5
Serial Receiver Overflow
= 4
Main Board EEPROM Error
= 2
Probe EEPROM (any) Error
= 1
Probe EEPROM Not Responding
= 8
Probe EEPROM Read Error
= 1
Probe EEPROM Checksum Error
= 4
Probe EEPROM Verification Error
= 2
Probe EEPROM Identification Error
= 1
Communications Syntax Error
= 4
Piezo Sensor Bad (voltage out of range)
= 2
Pirani Sensor Bad (voltage out of range)
= 1
High Setpoint Alarm (pressure exceeds setpoint)
= 4
Low Setpoint Alarm (pressure less than setpoint)
= 2
Gas# Changed (not the same as when unit last calibrated)
= 1
6.4 Installation of RS-232 Board
Materials needed for option board installation:
RS-232 option board
Connector panel for RS-232 option card
Screws (4)
Jackscrews (2)
Tools needed for option board installation:
Phillips screw driver
3/16 wrench (nut driver)
○○○
○○○○○○
○○○○○○
○○○○○○
○○○○○○○○○○○○○○
○○○○○○○○○○○○○○○○○○
○○○○○○○○○○○
○○○○○○○○○○○○○○○○○○
○○○○○○○○○○○○○○○○○○○○○