Teledyne HASTINGS 200 Series Instruction Manual Download Page 8

126B - PAGE 8

The standard inlet and outlet fittings for the HFM-229B Flowmeter are NPT thread in the size of the
pipe listed on the drawing on page 22, for sizes up to 3" laminar. On the 4", 6", and 8" laminars, flange
fittings are standard. Smooth tubulation is optional on all sizes.

It is suggested that all connections be checked for leaks after installation. This can be done by pressuriz-
ing the transducer (do not exceed 500 psig unless the transducer is specifically rated for higher pres-
sures) and applying a diluted soap solution to the flow connections.

2.5  Electrical Connections:

2.5  Electrical Connections:

2.5  Electrical Connections:

2.5  Electrical Connections:

2.5  Electrical Connections:

If a power supply was purchased from Hastings Instruments, installation will consist of connecting the
cable, purchased separately from the power supply, from the rear of the supply to the top of the
flowmeter. If a Hastings supply was not purchased, follow the instructions below to connect up the
flowmeter.

The power supply used must be capable of supplying ±15VDC at ±50mA . These voltages must be
referenced to a common ground.

Refer to Fig. 2.1. Connect -15VDC to pin 9 and +15VDC to pin 11.     Pin 5 is common and must be
connected to the common connection at the power supply. Pin 7 is the case ground and it should be
connected to the cable shield if available, and to AC ground of the power supply. Pin 6 is the output
signal from the flowmeter. This output will be 0-5VDC, 5VDC being 100% of rated or full flow.

2.6  Operation:

2.6  Operation:

2.6  Operation:

2.6  Operation:

2.6  Operation:

2.6.1  P

2.6.1  P

2.6.1  P

2.6.1  P

2.6.1  Pooooow

w

w

w

wer Supply Opera

er Supply Opera

er Supply Opera

er Supply Opera

er Supply Operation

tion

tion

tion

tion

Do not connect transducers while the power supply is energized. The display may read either percent of
full scale or actual flow rate, depending on the power supply purchased.  To read flow rate, turn display
switch to desired channel.

2.6.2  

2.6.2  

2.6.2  

2.6.2  

2.6.2  Ambient 

Ambient 

Ambient 

Ambient 

Ambient T

T

T

T

Tempera

empera

empera

empera

emperature

ture

ture

ture

ture

In order to maintain the accuracy of the flowmeter with changes in ambient temperature, it is necessary
to keep the temperature of the transducer between 0

o

C and 50

o

C. Since some of the temperature shift

results in a slight zero offset, better results are obtained if the flowmeter is re-zeroed at the operating
temperature. The flowmeter calibration may change by a factor up to 0.1%/

o

C. Hastings Mass

Flowmeters are for 

GAS

 flow, so 

DO NOT

 let the temperature and/or pressure of the gas

reach a point that would cause the gas to change to a liquid state, or erroneous indications will result.

2.6.3  Zero Check

2.6.3  Zero Check

2.6.3  Zero Check

2.6.3  Zero Check

2.6.3  Zero Check

Turn the power supply “ON”. Allow the flowmeter 10 minutes to warm up. Stop all flow through the
transducer and check electrical zero.

CAUTION:  Do not assume that all metering valves will completely shut off flow. Even a slight leakage
through a valve will cause an indication on the meter which will falsely appear to be a zero shift.

If necessary, adjust the “ZERO” potentiometer, located on the lower inlet side of the transducer, until
the meter indicates zero. This zero should be checked periodically during normal operation.

2.7  Range Changing:

2.7  Range Changing:

2.7  Range Changing:

2.7  Range Changing:

2.7  Range Changing:

The range of the flowmeter can be changed in the field if recalibration facilities are available. The
instructions to change the flow range can be found in Section 4.5. In order to change the range of a
Model HFM-230 Flowmeter, a new laminar flow element must be purchased from the factory.

Summary of Contents for HASTINGS 200 Series

Page 1: ...00 SERIES HASTINGS 200 SERIES HASTINGS 200 SERIES F F F F FAST RESPONSE AST RESPONSE AST RESPONSE AST RESPONSE AST RESPONSE MASS FLOWMETERS MASS FLOWMETERS MASS FLOWMETERS MASS FLOWMETERS MASS FLOWMETERS INSTRUCTION INSTRUMENTS TELEDYNE HASTINGS INSTRUCTION MANUAL ...

Page 2: ... should incorporate immediately into the manual Addenda are numbered sequentially When a new revision is created all addenda associated with the previous revision of the manual are incorporated into the new revision of the manual Each new revision includes a revised copy of this print history page Revision B Document Number 126 082000 August 2000 Revision C Document Number 126 082005 August 2005 H...

Page 3: ...4 Mechanical Connections 7 2 5 Electrical Connections 8 2 6 Operation 8 2 6 1 Power Supply Operation 8 2 6 2 AmbientTemperature 8 2 6 3 Zero Check 8 2 7 Range Changing 8 SECTION 3 0 THEORY OF OPERATION 9 3 1 Overall Functional Description 9 3 2 Sensor 9 3 3 Electronics 10 3 4 Shunt 10 SECTION 4 0 MAINTENANCE 13 4 1 Introduction 13 4 2 Authorized Maintenance 13 4 3 Adjustments 13 4 3 1 Calibration ...

Page 4: ... 15 SECTION 5 0 REPLACEMENT PARTS 17 SECTION 6 0 WARRANTY AND REPAIR 19 SECTION 7 0 DIAGRAMS AND DRAWINGS 21 AssemblyDiagram PC810FastReponse 22 Schematic PC 810 Fast Response 23 HFM 200 Outline Drawng 24 HFM 201 Outline Drawing 25 HFM 200B Outline Drawing 26 Laminar Flow Element Outline Drawing 27 ...

Page 5: ... a removable replaceable sensor which virtually eliminates long down time due to clogging the most common cause of failure in the industry Each flowmeter has a shunt which can be quickly and easily adjusted in the field to different ranges however recalibration is normally required A 100 micron filter is located upstream of the shunt and sensor to eliminate most of the larger impurities in the sys...

Page 6: ... 3 2 1 3 2 1 3 2 Alar Alar Alar Alar Alarms ms ms ms ms The Model AL 1 Flow Alarm is available as an attachment suitable for use with any 0 5 00VDC input signal Calibrated digital dial precision pots determine the low and high set points as a percent of full scale This permits use with any range flowmeter or other instrument having a 0 5 00VDC linear output signal Control action is within 0 2 of s...

Page 7: ...urge suppressers are recommended to prevent power line spikes from feeding through to the instruments These power requirements are satisfied by the Hastings Power Supply described in Section 1 4 Accessories 2 3 Output 2 3 Output 2 3 Output 2 3 Output 2 3 Output V V V V Voltage oltage oltage oltage oltage The output of the flowmeter is a 0 5 00VDC signal proportional to the flow rate The output is ...

Page 8: ...t transducers while the power supply is energized The display may read either percent of full scale or actual flow rate depending on the power supply purchased To read flow rate turn display switch to desired channel 2 6 2 2 6 2 2 6 2 2 6 2 2 6 2 Ambient Ambient Ambient Ambient Ambient T T T T Tempera empera empera empera emperature ture ture ture ture In order to maintain the accuracy of the flow...

Page 9: ... sensor and the shunt is always laminar The circuit board amplifies the sensor output to 5 00VDC and sends it to the power supply to display All of these components working together result in a fast stable flowmeter 3 2 Sensor 3 2 Sensor 3 2 Sensor 3 2 Sensor 3 2 Sensor The Hastings Model HFM Mass Flowmeter operates on a unique thermal electric principle whereby a metallic capillary tube is heated...

Page 10: ...rate ofchange of the input signal and adding it to the amplified input signal The elec tronics are adjusted to provide response times of less than two seconds to 98 of a 0 100 step change with less than 2 overshoot The electronics also maintain the ouput signal above 5 00VDC when the flowrate exceeds the full scale flow of the flowmeter NO NO NO NO NOTE TE TE TE TE The output signal will continue ...

Page 11: ...is adjustable from 0 50 slpm to 0 500 slpm The HFM 229B Series uses an external laminar flow element instead of an internal shunt because the necessary size for the element exceeds the body size of the flowmeter This laminar flow element is made of a corrugated and fused stainless steel ribbon similar to the one used in the Model HFM 230 Fig 3 7 Fig 3 7 Fig 3 7 Fig 3 7 Fig 3 7 ...

Page 12: ...126B PAGE 12 ...

Page 13: ... 4 4 4 4 4 Adjust SPAN pot until the flowmeter indicates full scale flow 5 000VDC NOTE Perform this step only if a calibrated reference flowmeter is avail able 5 5 5 5 5 Record flowmeter and flow reference outputs for flow rates of 20 40 60 80 and 100 4 3 2 Response 4 3 2 Response 4 3 2 Response 4 3 2 Response 4 3 2 Response Time Time Time Time Time Adjustments Adjustments Adjustments Adjustments ...

Page 14: ... 5 Range Changes 4 5 Range Changes Remove end cap and shunt assembly per Section 4 4 above The 0 10 sccm to 0 300 sccm range shunt has tubes while the medium range shunt is built from a coil of corrugated stainless steel foil The medium range shunt has a disk assembly with two large one medium and one small hole To change ranges the disc on the inlet of the shunt should be loosened turned and reti...

Page 15: ... and tighten down the three screws evenly Replacement of sensor will require recalibration per Section 4 3 1 4 8 4 8 4 8 4 8 4 8 T T T T Troubleshooting roubleshooting roubleshooting roubleshooting roubleshooting SYMPT SYMPT SYMPT SYMPT SYMPTOM OM OM OM OM Output reads 40 of flow with no flow Zero pot has no effect CA CA CA CA CAUSE USE USE USE USE Power supply locked up or shorted out A A A A ACT...

Page 16: ...an 0 00VDC with no flow or there is a small flow when flowmeter reads 0 00VDC CA CA CA CA CAUSE USE USE USE USE ZERO potentiometer is out of adjustment A A A A ACTION CTION CTION CTION CTION Shut off all flow Adjust ZERO potentiometer until output reads 0 00VDC SYMPTOM SYMPTOM SYMPTOM SYMPTOM SYMPTOM Flowmeter out of calibration and non linear CAUSE CAUSE CAUSE CAUSE CAUSE Leaks in gas inlet or ou...

Page 17: ...4 0 Flange 0 3000 SLPM 65 140K LFE Model LS 6F 6 0 Flange 0 6000 SLPM 65 140L LFE Model LS 8F 8 0 Flange 0 15 000 SLPM 39 02 003 SMALL FILTER DISC HFM 229 HFM 229B AND HFC 202 39 02 002 LARGE FILTER DISC HFM 230AND HFC 203 81 105 SENSOR MODULE ALL MODELS 65 595 FAST RESPONSE HFM ELECTRONICS CARD D CONNECTOR 81 115 VITON O RING KIT FOR HFM 200 and 200B 81 116 VITON O RING KIT FOR HFM 201 81 146 KAL...

Page 18: ...126B PAGE 18 ...

Page 19: ...nstruments determines in good faith that it is defective in materials or workmanship during the warranty period Defective instruments should be returned to Hastings Instruments together with a written statement of the problem and a Return MaterialAuthorization RMA number Please consult the factory for your RMA number before returning any product for repair Non W Non W Non W Non W Non War ar ar ar ...

Page 20: ...126B PAGE 20 ...

Page 21: ...126B PAGE 21 7 0 Diagrams and Drawings ...

Page 22: ...126B PAGE 22 PC 810Assembly InformationfromHastingsdrawing30837rev F ...

Page 23: ...126B PAGE 23 PC 810Schematic InformationfromHastingsdrawing30837rev F ...

Page 24: ...126B PAGE 24 ...

Page 25: ...126B PAGE 25 ...

Page 26: ...126B PAGE 26 ...

Page 27: ...126B PAGE 27 ...

Page 28: ...126B PAGE 28 ...

Reviews: