126B - PAGE 16
SYMPTOM:
SYMPTOM:
SYMPTOM:
SYMPTOM:
SYMPTOM: Sensor has proper resistance readings, but little or no output with flow.
CAUSE:
CAUSE:
CAUSE:
CAUSE:
CAUSE: Plugged sensor.
A
A
A
A
ACTION:
CTION:
CTION:
CTION:
CTION: Shut off gas supply and power supply. Remove cover and PC board from unit. Remove
sensor assembly and examine. If sensor has evidence of plugging, clean or replace as applicable.
SYMPT
SYMPT
SYMPT
SYMPT
SYMPTOM:
OM:
OM:
OM:
OM: Flowmeter reads other than 0.00 VDC with no flow, or there is a small flow when
flowmeter reads 0.00 VDC.
CA
CA
CA
CA
CAUSE:
USE:
USE:
USE:
USE: ZERO potentiometer is out of adjustment.
A
A
A
A
ACTION:
CTION:
CTION:
CTION:
CTION: Shut off all flow. Adjust ZERO potentiometer until output reads 0.00 VDC.
SYMPTOM:
SYMPTOM:
SYMPTOM:
SYMPTOM:
SYMPTOM: Flowmeter out of calibration and non-linear.
CAUSE:
CAUSE:
CAUSE:
CAUSE:
CAUSE: Leaks in gas inlet or outlet fittings.
A
A
A
A
ACTION:
CTION:
CTION:
CTION:
CTION: Check all fittings for leaks by placing soap solution on all fittings between gas supply and
final destination of gas. Check flowmeter for leaks. Replace “O” rings if required or recalibrate as
necessary.
Summary of Contents for HASTINGS 200 Series
Page 12: ...126B PAGE 12 ...
Page 18: ...126B PAGE 18 ...
Page 20: ...126B PAGE 20 ...
Page 21: ...126B PAGE 21 7 0 Diagrams and Drawings ...
Page 22: ...126B PAGE 22 PC 810Assembly InformationfromHastingsdrawing30837rev F ...
Page 23: ...126B PAGE 23 PC 810Schematic InformationfromHastingsdrawing30837rev F ...
Page 24: ...126B PAGE 24 ...
Page 25: ...126B PAGE 25 ...
Page 26: ...126B PAGE 26 ...
Page 27: ...126B PAGE 27 ...
Page 28: ...126B PAGE 28 ...