
Calibration Procedures Teledyne API – Model T300/T300M CO Analyzer
Teledyne
Analytical
Instruments
199
9.4.1.
SETUP
Æ
ACAL: PROGRAMMING AND AUTO CAL
SEQUENCE
Note
If at any time an illegal entry is selected, (for example: Delta Days > 366)
the ENTR label will disappear from the control button.
To program the example sequence shown in Table 9-4, press:
CONTINUE NEXT PAGE
With
STARTING TIME
SAMPLE RANGE = 50.0 PPM CO=XX.XX
< TST TST > CAL CALZ CZLS
SETUP
SETUP X.X
CFG
ACAL
DAS RNGE PASS CLK MORE EXIT
SETUP X.X
SEQ 1) DISABLED
NEXT
MODE
EXIT
SETUP X.X
SEQ 2) DISABLED
PREV NEXT
MODE
EXIT
SETUP X.X
MODE: DISABLED
NEXT
ENTR EXIT
SETUP X.X
MODE: ZERO
PREV
NEXT
ENTR EXIT
SETUP X.X
SEQ 2) ZERO–SPAN, 1:00:00
PREV NEXT
MODE
SET
EXIT
SETUP X.X
TIMER ENABLE: ON
SET>
EDIT
EXIT
SETUP X.X
STARTING DATE: 01–JAN–07
<SET SET>
EDIT
EXIT
SETUP X.X
STARTING DATE: 01–JAN–02
0
4 SEP
0
8
ENTR
EXIT
Toggle buttons to set
Day, Month & Year:
Format : DD-MON-YY
SETUP X.X
MODE: ZERO–SPAN
PREV NEXT
ENTR
EXIT