5.7.
SETUP
Æ
COMM: COMMUNICATIONS PORTS
This section introduces the communications setup menu; Section 6 provides the setup
instructions and operation information. Press SETUP>ENTR>MORE>COMM to arrive
at the communications menu.
5.7.1.
ID (MACHINE IDENTIFICATION)
Press ID to display and/or change the Machine ID, which must be changed to a unique
identifier (number) when more than one instrument of the same model is used:
•
in an RS-232 multidrop configuration (Sections 3.3.1.9 and
Error! Reference
source not found.
)
•
on the same Ethernet LAN (Section 6.5)
The default ID is the same as the model number; for the Model T100, the ID is 0100.
Press any button(s) in the MACHINE ID menu (Figure 5-2) until the Machine ID
Parameter field displays the desired identifier.
SETUP X.X
COMMUNICATIONS MENU
ID
INET
COM1
COM2
EXIT
SETUP X.
MACHINE ID: 300 ID
0 3 0 0
ENTR EXIT
Toggle
to cycle
through the available
character set:
0-9
ENTR accepts the new
settings
EXIT ignores the new
settings
Figure 5-2:
COMM– Machine ID
The ID can be any 4-digit number and can also be used to identify analyzers in any
number of ways (e.g. location numbers, company asset number, etc.)
5.7.2.
INET (ETHERNET)
Use SETUP>COMM>INET to configure Ethernet communications, whether manually
or via DHCP. Please see Section 6.5 for configuration details.
5.7.3.
COM1 AND COM2 (MODE, BAUD RATE AND TEST PORT)
Use the SETUP>COMM>COM1[COM2] menus to:
•
configure communication modes (Section 6.2.1
•
view/set the baud rate (Section 6.2.2)
•
test the connections of the com ports (Section 6.2.3).
Configuring COM1 or COM2 requires setting the DCE DTE switch on the rear panel.
Section 6.1 provides DCE DTE information.