7.1.11.
HOLDOFF FEATURE
The DAS
HOLDOFF
feature prevents data collection during calibration operations and
at certain times when the quality of the analyzer’s CO measurements may not be certain
(e.g. while the instrument is warming up). In this case, the length of time that the
HOLDOFF
feature is active is determined by the value of the internal variable (
VARS
),
DAS_HOLDOFF.
To set the length of the DAS_HOLDOFF period, go to the SETUP>MORE>VARS
menu (Section 5.8) and EDIT the “0) DAS_HOLD_OFF…” parameter.
To enable or disable the
HOLDOFF
feature for an individual channel, follow the
instruction shown in Section 7.1.4 then press: