All other interfaces operate on the data processing clock of the FPGA at 250 MHz. This clock is referred
to as the Data Clock.
See
for more details on the clock system.
1.2.6
Analog signal range
The analog signal range (
ACTUAL_ANALOG_RANGE
) is symmetrical around zero. The value of the
analog signal range it is depending on the gain setting. With for example a range of, 500 mV
pp
, the ana-
log input signal can vary from –250 mV to +250 mV and the range can be moved from [–0 mV
+500 mV] to [–500 mV +0 mV] by the DC-offset feature,
.
The maximum digital code 2^15 represents an analog signal with a level
ACTUAL_ANALOG_RANGE / 2
at the input. A specific analog signal
ANALOG_LEVEL
will then be represented by the following digital
code:
DIGITAL_CODE_LEVEL = ANALOG_LEVEL / ( ACTUAL_ANALOG_RANGE / 2 ) * 2^15
(1)
A specific code
DIGITAL_CODE_LEVEL
then represent the analog level as:
ANALOG_LEVEL = ( DIGITAL_CODE_LEVEL / 2^15 ) * ( ACTUAL_ANALOG_RANGE / 2)
(2)