2.3 Installing Bottle Kits
When installing bottles or a new kit, keep these
guidelines in mind:
• Do not mix glass and plastic bottles
together.
• Install all bottles to insure that they
remain in place.
• Align the bottles correctly. If the bottles
are misaligned, the sampler may miss
the bottle mouth, or deposit samples in
the “wrong” bottle.
• Each time you change bottle sizes, you
must adjust the distributor arm and
install a discharge tube of the correct
length.
• Each time you install a bottle kit, check
the program settings for the number of
bottles and bottle volume. Also rotate the
distributor arm to each bottle position to
ensure the arm is properly positioned
over each bottle. When rotating the
distributor arm use the controller
software.
Do not manually rotate the
distributor arm.
• When using the refrigerator you may
need to readjust the stops and ramps,
and use a different distributor shaft
extension.