4-6
4 Operations
4 Operations
4 Operations
4 Operations
4 Operations
TELEDYNE ANALYTICAL INSTRUMENTS
1. At the main Analyze mode screen press ZERO.
2. Press ACK to acknowledge that a low O
2
gas has al-
ready been connected to the sample port. The sample
gas O
2
level should be close to zero to preserve the life
of the scrubber (if one is installed).
3. Set the flow rate to 2 SCFH using the knob on the sys-
tem front panel and press ACK.
4. Press AUTO to let the system determine the settling
time before spanning or press MAN for manual.
5. If MAN was chosen in step 4, watch the logger or the
O
2
value on the screen to assess when the output is con-
stant, and then press ACK.
6. After step 5, the system will automatically span to the
user-entered span value. If the zeroing is successful, the
main Analyze mode menu will appear as in step 10.
Calibration Zeroing
Calibration Zeroing
Calibration Zeroing
Calibration Zeroing
Calibration Zeroing