2150EX Area Velocity Flow Module
Section 2 Preparation and Installation
2-4
2.3 Site Examples
Figures 2-1, 2-2, and 2-3 illustrate typical round-pipe sites. Key
items are called out in the illustration and explained below.
Figures 2-1 and 2-2 represent
portable installations
. For
details about portable installations, see Section 2.4.
Figure 2-3 represents a
permanent installation
. For details
about permanent installations, see Section 2.5.
The
computer running Flowlink
(Figures 2-1 and 2-3) or the
2101 Field Wizard module
(Figure 2-2) should be located
outside the potentially explosive atmosphere. The computer and
modules communicate with the 2150EX module.
The
2150EX area velocity flow module
measures and stores
the stream data. In portable installations (Figures 2-1 and 2-2),
it is attached to a
2191EX or 2196EX battery module
, which
supplies power to the module.
As described in Section 2.6.1, the
EX network cable
connects to
the top of the 2150EX stack and extends to the interface of the
safe and hazardous areas.
As described in Section 2.6.2, an
RS232EX isolator
cable
con-
nects the computer and the site. The cable supports the data
transfers between the two, and is connected to an EX Network
Cable
connected to the top of the 2150EX module.
As described in Section 2.6.3, an
RS485EX isolator cable
con-
nects the site with a Field Wizard or other network device. The
cable supports the data transfers between the two, and is con-
nected to an EX Network Cable, connected to the top of the
2150EX module.
In permanent installations (Figure 2-3), the 2150EX is connected
via a network interface cable, usually through conduit, to the
2194EX network module
, located in the safe area, which
serves as both power supply and network or PC connection.
The
AV2150EX sensor cable
must be routed carefully without
kinks, coils, or sharp bends, but may be snake-looped and tied.
Any excess cable must be kept out of the channel to prevent
debris from collecting.
The
Mounting Ring
holds the AV2150EX sensor in place.
The
AV2150EX sensor
is positioned in the flow stream to
measure liquid level and velocity.
Summary of Contents for 2150EX
Page 2: ......
Page 4: ......
Page 26: ...2150EX Area Velocity Flow Module Section 1 Introduction 1 12 ...
Page 78: ...2150EX Area Velocity Flow Module Section 4 Modbus Protocol 4 10 ...
Page 90: ...2150EX Area Velocity Flow Module Section 5 Maintenance 5 12 ...
Page 92: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 2 ...
Page 93: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 3 ...
Page 94: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 4 ...
Page 95: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 5 ...
Page 96: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 6 ...
Page 98: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 8 602002331 A ...
Page 99: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 9 ...
Page 100: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 10 ...
Page 101: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 11 ...
Page 102: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 12 ...
Page 103: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 13 ...
Page 104: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 14 ...
Page 105: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 15 ...
Page 106: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 16 ...
Page 107: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 17 ...
Page 108: ...2150EX Area Velocity Flow Module Appendix A Replacement Parts A 18 ...
Page 134: ...Compliance Statements ...
Page 136: ......
Page 138: ......
Page 140: ......
Page 142: ......
Page 144: ...arranty ...
Page 146: ......