3 Operation
Model 2010A
3-2: Part II
Teledyne Analytical Instruments
2
15 V Failure
3
Both Failed
Analog
0
OK
1
DAC A (0–1 V Concentration)
2
DAC B (0–1 V Range ID)
3
Both Failed
Preamp
0
OK
1
Zero too high
2
Amplifier output doesn't match test input
3
Both Failed
The results screen alternates for a time with:
Press Any Key
To Continue...
Then the analyzer returns to the initial System screen.
3.2
Cell Failure Checks
Cell failure is covered in detail in Part I:
Control Units
, section 4.4.1.3,
Cell Failure
. Cell replacement is covered Part II:
Analysis Units
chapter 5,
Maintenance
.