Teledyne Analytical Instruments
Thermal Conductivity Analyzer
Part I: Control Unit
Part I 3-
(Reset by pressing
I/o
button to remove power. Then
press
I/o
again and any other button EXCEPT
Sys-
tem
to resume
.
Further detail can be found in chapter 4, section 4-5.
Figure 3-: Types of Relay Contacts
3.3.5 Digital Remote Cal Inputs
Accept 0 V (off) or 24
V dc (on) inputs for remote control of calibra-
tion. (See
Remote Calibration Protocol
below.)
Zero:
Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the
Zero
mode. Either side may be
grounded at the source of the signal. A synchronous signal
must open and close the external gas control valves appro-
priately. See 3.3.9
Remote Probe Connector
. (With the –C
option, the internal valves operate automatically.)
Span:
Floating input. 5 to 24 V input across the + and – terminals
puts the analyzer into the
Span
mode. Either side may be
grounded at the source of the signal. A synchronous signal
must open and close the external gas control valves appro-
priately. See 3.3.9
Remote Probe Connector
. (With the –C
option, the internal valves operate automatically.)
Cal Contact:
This relay contact is closed while analyzer is spanning
and/or zeroing. (See
Remote Calibration Protocol
below.)
Remote Calibration Protocol:
To properly time the Digital Remote
Cal Inputs to the Model 2010A Analyzer, the customer's controller must
monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal
Inputs are being polled, and a zero or span command can be sent.
Summary of Contents for 2010A
Page 16: ...Teledyne Analytical Instruments 1 Part 1 Introduction Model 2010A ...
Page 22: ...2 6 Part I 2 Operational Theory Model 2010A Teledyne Analytical Instruments ...
Page 34: ...Teledyne Analytical Instruments 3 Installation Model 2010A 3 12 Part I ...
Page 70: ...5 Maintenance Model 2010A 5 6 Part I Teledyne Analytical Instruments ...
Page 86: ...1 4 Part II Teledyne Analytical Instruments 1 Introduction Model 2010A ...
Page 104: ...4 Maintenance Model 2010A 4 8 Part II Teledyne Analytical Instruments ...