User Manual
13
Table of Required Test Equipment
Description
Minimum Requirement
Example Equipment
Digital Oscilloscope
ProBus Interface
Windows-based
≥ 2 GHz for Bandwidth
Test
WaveRunner 8000
WavePro 7 Zi-A
Digital Multimeter
(DMM) with test probe
leads
4.5 digit
DC: 0.1% Accuracy
AC: 0.1% Accuracy
Keysight 34401A
Fluke 8842A-09
Function Generator
Sine Wave output
amplitude adjustable to
14.14 Vp-p (5 Vrms) into
1 MΩ at 70 Hz
Keysight 33120A
Stanford Research DS340
RF Signal Generator
Sine wave output 500
MHz, 1 GHz, 1.5 GHz
+6 dBm output
Anritsu MG3690C
Power Supply
0-12 V, settable to 10 mV
HP E3611A
BNC Coaxial Cable
(2)
Male to Male, 50 Ω,
36" Cable
Pomona 2249-C-36
Pomona 5697-36
BNC Tee Connector
Male to Dual Female
Pomona 3285
Calibration Fixture
0.1 mm square pin head
Teledyne LeCroy PCF200
Calibration Fixture
ProBus Extender Cable
Teledyne LeCroy
PROBUS-CF01
Terminator,
Precision BNC
50 Ω ± 0.05%
Teledyne LeCroy
TERM-CF01
Banana Plug Adapter
(2)
Female BNC to
Dual Banana Plug
Pomona 1269
BNC to Mini-grabber
BNC Male to
Mini-grabber Cable, 36”
Pomona 5187-C-36
2.54mm Sq. Pin Short
Pins connected to short
the probe inputs
Samtec TSW-102-07-G-S
SMA-BNC Adapter