LabMaster 10 Zi-A Getting Started Guide
11
SET UP
Powering On/Off
Acquisition modules are shipped with a
cable
clamp
attached to the AC power inlet. Use a Philips
screwdriver to loosen the screw on the side of the
clamp, connect the power cord, then tighten the
screw to hold the cable connector securely in place.
Plug all modules into grounded, AC power outlets.
With the system fully connected, press the
Power
button
on the front of the MCM-Zi-A to turn on the
system. The button will light to show the system is
operational.
NOTE:
If any cable connections are incorrect, the system will not power on.
CAUTION.
Do not power on or calibrate with a signal attached.
Press the Power button again or use or the
File > Shutdown
menu option
to execute a proper shut down process and preserve settings.
CAUTION
. Do not press and hold the Power button. This will
execute a hard shutdown, but may not preserve setups and data.
The Power button does not disconnect the LabMaster from the AC power
supply but puts it into "Standby" mode. Some “housekeeping” circuitry
continues to draw power. The only way to fully power down the system is
to unplug the power cords. We recommend unplugging the system if it will
remain unused for a long period of time.
CAUTION.
Do not change the Windows® Power setting to System
Standby or System Hibernate. Doing so may cause failure.
Probes and Adapters
Visit
teledynelecroy.com/probes
for probe specifications and
documentation
.
See the probe product page for LabMaster compatibility.
A variety of adapters are available to accommodate cables that have
different type interfaces than your acquisition modules.
• A 1.85 mm V-female to V-female barrel adapter is delivered with
≥ 50 GHz bandwidth acquisition modules. The male connector is
mounted onto the oscilloscope connector panel when delivered. The
barrel adapter must be manually attached before connecting cables.
• Two similar barrel adapters are included with 100 GHz acquisition
modules: a 1 mm female to 1 mm female adapter, and a 1 mm female
to 1.85 mm female adapter. Like the 1.85 mm adapter, they must be
manually attached before connecting cables. Use the 4ft•lb torque
wrench included standard.
• The optional L2.92A-PLINK and L2.92A-PBUS adapters enable you to
connect a range of ProLink or ProBus probe platforms to the 2.92mm
and LEMO interfaces.
See the
LabMaster Oscilloscope Systems Operator’s Manual
or LabMaster
Support for instructions on attaching adapters.