4

Mounting the Tray 

1

Place the tray on the base of the autosampler. 

Make sure the locating pins on the tray fully seat into the locating holes in the base.

Installing the Stirrer

When the autosampler is used for oils, install the stirrer paddle. When the autosampler is 

used for coolants, the stirrer paddle should be removed (or replaced with the optional 

coated coolants stirrer paddle).

1

Locate the oils stirrer paddle. 

 

F

IGURE 

4

    Stirrer Paddle for Oils

2

Press the stirrer paddle into the hole on the bottom of the stirrer motor. 

 

F

IGURE 

5

    Inserting the Stirrer Paddle

3

Place the stirrer motor into the larger hole in the probe/stirrer block on the Z-drive. 

 

F

IGURE 

6

    Placing the Stirrer Motor

Summary of Contents for Oils 7400 Series

Page 1: ...Quick Installation Guide Manual Part Number 32 0356 048 Rev0 July 18 2018 2018 Teledyne Technologies Incorporated All rights reserved Printed in USA Oils 7400 Series Homogenizing Autosampler...

Page 2: ...tions on installing the software and information on using and caring for the system System Components FIGURE 1 Oils 7400 Autosampler Front View FIGURE 2 Z Drive Assembly Right Side View Autosampler He...

Page 3: ...r Lifting should be done with a person situated on either side of the instrument Lifting without assistance may cause injury NOTE Keep the original packaging for use in case the product ever needs to...

Page 4: ...When the autosampler is used for coolants the stirrer paddle should be removed or replaced with the optional coated coolants stirrer paddle 1 Locate the oils stirrer paddle FIGURE 4 Stirrer Paddle for...

Page 5: ...propriate probe for your application 2 Locate the two thumbscrews which are supplied with the probe 3 Guide the probe straight down through the smaller hole in the Z drive 4 Adjust the height of the p...

Page 6: ...ion will be recycled by returning it to the rinse solution bottle If necessary rinse solution can be pulled from a fresh bottle and used solution drained to a waste container Remember to label the was...

Page 7: ...altic pump 1 Connect up to 1 8 meters of tubing from the outlet fitting of the pump the lower fitting of the channel closest to the autosampler to the waste container 3 33 FIGURE 10 Connecting the Was...

Page 8: ...rinse station and the Oils pump should be connected to the green side of the rinse station FIGURE 12 Connecting the Rinse Station Drain Oils Side Is Shown 5 Connect the second piece of tubing from ste...

Page 9: ...expectedly and cause an injury 1 Before loading or unloading any sample vial racks on the sample tray park the sampling arm and probe in the home position The home position is the initial position at...

Page 10: ...power supply Ensure that you position the autosampler so that the location where the power supply cord plugs into it is easily accessible is not blocked and it can be quickly disconnected if needed In...

Page 11: ...OTE Use either a USB cable or a serial cable but not both The serial port will not function when a USB cable is connected 1 Power on both the computer and the autosampler 2 Plug one end of the cable i...

Page 12: ...peristaltic pump on the back of the autosampler Visually verify that the tubing is directed to the correct rinse supply and waste containers 3 Replace the sample probe with the appropriate version 4...