background image

12

Switching Between Coolant and Oil Samples

1

Move the rinse station to the appropriate position.  The status LED on the front of the 

autosampler will show green for the oils position or blue for the coolants position.

      

 

F

IGURE 

16

    Rinse station in coolants (left) and oils (right) position.

2

Moving the rinse station will automatically switch to the appropriate peristaltic pump on the 

back of the autosampler. Visually verify that the tubing is directed to the correct rinse supply 

and waste containers. 

3

Replace the sample probe with the appropriate version.

4

Replace the stirring paddle with the appropriate version. For oils, use the oils paddle. For 

coolants, remove the paddle or use the optional coolants paddle (identified by a black band).

5

If using an 

ASXP

RESS 

P

LUS

 Rapid Sample Introduction System, either replace the valve and 

tubing, or use a second valve/pump module which has been set up for oils or coolants.  

Summary of Contents for Oils 7400 Series

Page 1: ...Quick Installation Guide Manual Part Number 32 0356 048 Rev0 July 18 2018 2018 Teledyne Technologies Incorporated All rights reserved Printed in USA Oils 7400 Series Homogenizing Autosampler...

Page 2: ...tions on installing the software and information on using and caring for the system System Components FIGURE 1 Oils 7400 Autosampler Front View FIGURE 2 Z Drive Assembly Right Side View Autosampler He...

Page 3: ...r Lifting should be done with a person situated on either side of the instrument Lifting without assistance may cause injury NOTE Keep the original packaging for use in case the product ever needs to...

Page 4: ...When the autosampler is used for coolants the stirrer paddle should be removed or replaced with the optional coated coolants stirrer paddle 1 Locate the oils stirrer paddle FIGURE 4 Stirrer Paddle for...

Page 5: ...propriate probe for your application 2 Locate the two thumbscrews which are supplied with the probe 3 Guide the probe straight down through the smaller hole in the Z drive 4 Adjust the height of the p...

Page 6: ...ion will be recycled by returning it to the rinse solution bottle If necessary rinse solution can be pulled from a fresh bottle and used solution drained to a waste container Remember to label the was...

Page 7: ...altic pump 1 Connect up to 1 8 meters of tubing from the outlet fitting of the pump the lower fitting of the channel closest to the autosampler to the waste container 3 33 FIGURE 10 Connecting the Was...

Page 8: ...rinse station and the Oils pump should be connected to the green side of the rinse station FIGURE 12 Connecting the Rinse Station Drain Oils Side Is Shown 5 Connect the second piece of tubing from ste...

Page 9: ...expectedly and cause an injury 1 Before loading or unloading any sample vial racks on the sample tray park the sampling arm and probe in the home position The home position is the initial position at...

Page 10: ...power supply Ensure that you position the autosampler so that the location where the power supply cord plugs into it is easily accessible is not blocked and it can be quickly disconnected if needed In...

Page 11: ...OTE Use either a USB cable or a serial cable but not both The serial port will not function when a USB cable is connected 1 Power on both the computer and the autosampler 2 Plug one end of the cable i...

Page 12: ...peristaltic pump on the back of the autosampler Visually verify that the tubing is directed to the correct rinse supply and waste containers 3 Replace the sample probe with the appropriate version 4...

Reviews: