15
4. T100U OPERATING INSTRUCTIONS
Note
For the most part the operation instructions for the T100U are the same
as those described in the T100 Manual with the exception that there are
additional test parameters and setup procedures.
4.1 ADDITIONAL TEST PARAMETERS
Please see Section 3.3 of this addendum for details on the additional test parameters.
4.2 STBL MENU: SETUP FOR THE THREE STABILITY
FUNCTIONS
There is an additional submenu on the Secondary Setup Menu in the T100U’s software.
The STBL menu, see Figure 4-1 allows the user to modify the settings for the three
stability calculations that are displayed as Test Functions on the front panel and are
available via the serial data port and that can be logged with the DAS. Test Function
STABIL, CONC1 in the menus below, is equivalent to the standard T100 STABIL
function and is useful when conducting calibrations and other operations in which the
operator has limited time to view the display. CONC2, which is the same as the
STABIL2 test function has been configured so that it calculates stability in the same way
as required by the EPA. STABIL2 is useful when comparing instruments against the
EPA standard but is very slow and difficult to use for calibration or other activities
where the operator must wait for analyzer to settle to the desired value. UVLAMP is
the same as UV STAB on the front panel is a diagnostic that can be used to understand
the stability of the UV lamp.
06840B DCN 6201
Summary of Contents for T100U
Page 2: ......
Page 42: ...A 2 This page intentionally left blank 06840B DCN6201 ...
Page 44: ...This page intentionally left blank B 2 06840B DCN6201 ...
Page 52: ...D 2 This page intentionally left blank 06840B DCN6201 ...
Page 56: ...This page intentionally left blank D 6 06840B DCN6201 ...
Page 57: ...D 7 06840B DCN6201 ...