Percent Oxygen Analyzer
Appendix
A–1
Teledyne Analytical Instruments
Appendix
Specifications
TBE/AI Sales Order Number:
Instrument Model Number: 311PC
Instrument Serial Number:
Micro-fuel Cell Class: B-1
Accuracy: ±2% of scale at constant temperature;
±5% of full scale over the operating
temperature range.
Operating Temperature Range: 30
o
F to 125
o
F
Response and Recovery: At the specified flowrate (1–2 l/min)
90% in 10 seconds.
Ranges of Analysis: 0–25% Oxygen
0–10% Oxygen
0–5% Oxygen
0–2.5% Oxygen
0–1% Oxygen
Recommended Span Gas: Atmospheric air, or span gas 70–90%
of full scale on range of interest.