4 Operation
4 Operation
4 Operation
4 Operation
4 Operation
Model 3000PA
4-12
Teledyne Analytical Instruments
4.4.2.2
Manual Mode Spanning
Press
Span
to start the
Span
function. The screen that appears allows
you to select whether the span calibration is to be performed automatically or
manually.
Span: Settling:MAN
<ENT> For Next
Use the
∆∇
keys to toggle between
AUTO
and
MAN
span settling.
Stop when
MAN
appears, blinking, on the display. Press
Enter
to move to
the next screen.
Span Val: 20.90
<ENT>Span <UP>Mod #
Press
∆
(
<UP>
) to permit modification (
Mod #
) of span value.
Use the arrow keys to enter the oxygen concentration of the span gas
you are using (20.90 if you are using air). The
< >
arrows chose the digit,
and the
∆∇
arrows choose the value of the digit.
Press
Enter
to enter the span value into the system and begin the span
calibration.
Once the span has begun, the microprocessor samples the output at a
predetermined rate. It calculates the difference between successive samplings
and displays this difference as
Slope
on the screen. It takes several seconds
for the first
Slope
value to display.
Slope
indicates rate of change of the
Span
reading. It is a sensitive indicator of stability.
#### % Span
Slope=#### ppm/s
When the
Span
value displayed on the screen is sufficiently stable,
press
Enter
. (Generally, when the
Span
reading changes by 1 % or less of
the full scale of the range being calibrated for a period of five minutes it is
sufficiently stable.) Once
Enter
is pressed, the
Span
reading changes to the
correct value. The instrument then automatically enters the
Analyze
func-
tion.
4.5
The
Alarms Function
The Model 3000PA is equipped with 2 fully adjustable concentration
alarms and a system failure alarm. Each alarm has a relay with a set of form
“C" contacts rated for 3 amperes resistive load at 250 V ac. See Figure in
Chapter 3, Installation and/or the Interconnection Diagram included at the
back of this manual for relay terminal connections.
Summary of Contents for 3000PA-EU
Page 2: ......
Page 3: ......
Page 4: ......
Page 56: ...Percent Oxygen Analyzer Operation 4 4 19 Teledyne Analytical Instruments...
Page 77: ...A 10 Appendix Appendix Appendix Appendix Appendix Model 3000PA Teledyne Analytical Instruments...
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......