
Percent Oxygen Analyzer
Operation 4
4-5
Teledyne Analytical Instruments
4.3.2 Setting up an Auto-Cal
When proper automatic valving is connected (see chapter 3, installa-
tion), the Analyzer can cycle itself through a sequence of steps that automati-
cally zero and span the instrument.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3000PA is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.
To setup an
AutoCal
cycle:
Choose
System
from the Function buttons. The VFD will display five
subfunctions.
Contrast AutoCal
PSWD Logout More
Use < > arrows to blink
AutoCal
, and press
Enter
. A new screen for
Span/Zero
set appears.
Span OFF Nxt: 0d 0h
Zero OFF Nxt: 0d 0h
Press < > arrows to blink
Span
(or
Zero
), then press
Enter
again. (You
won’t be able to set
OFF
to
ON
if a zero interval is entered.) A
Span
Every
... (or
Zero Every
...) screen appears.
Span Every 0 d
Start 0 h from now
Use
∆∆∆∆∆∇
arrows to set an interval value, then use < > arrows to move to
the start-time value. Use
∆∆∆∆∆∇
arrows to set a start-time value.
To turn
ON
the
Span
and/or
Zero
cycles (to activate
Auto-Cal
): Press
System
again, choose
AutoCal
,
and press
Enter
again. When the
Span/
Zero
screen appears, use the < > arrows to blink the
Span
(or
Zero
)
OFF
/
ON
field. Use
∆∆∆∆∆∇
arrows to set the
OFF
/
ON
field to
ON
. You can now turn
these fields
ON
because there is a nonzero span interval defined.
4.3.3 Password Protection
If a password is assigned, then setting the following system parameters
can be done only after the password is entered: span and zero settings,
alarm setpoints, analysis range definitions, switching between autoranging
and manual override, setting up an auto-cal, and assigning a new password.
However, the instrument can still be used for analysis or for initiating a self-
test without entering the password.
Contrast Function is
DISABLED
(Refer to Section 1.6)
Summary of Contents for 3000PA-EU
Page 2: ......
Page 3: ......
Page 4: ......
Page 56: ...Percent Oxygen Analyzer Operation 4 4 19 Teledyne Analytical Instruments...
Page 77: ...A 10 Appendix Appendix Appendix Appendix Appendix Model 3000PA Teledyne Analytical Instruments...
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......