4 Operation
4 Operation
4 Operation
4 Operation
4 Operation
Model 3000PA
4-2
Teledyne Analytical Instruments
4.2
Using the Data Entry and Function
Buttons
Data Entry Buttons: T
he
< >
a
rrow buttons select options from the
menu currently displayed on theVFD screen. The selected option blinks.
When the selected option includes a modifiable item, the
∆∆∆∆∆∇
arrow
buttons can be used to increment or decrement that modifiable item.
The
Enter
button is used to accept any new entries on the VFD screen.
The
Escape
button is used to abort any new entries on the VFD screen that
are not yet accepted by use of the
Enter
button.
Figure 4-1 shows the hierarchy of functions available to the operator via
the function buttons. The six function buttons on the analyzer are:
•
Analyze.
This is the normal operating mode. The analyzer
monitors the oxygen content of the sample, displays the percent
of oxygen, and warns of any alarm conditions.
•
System.
The system function consists of six subfunctions that
regulate the internal operations of the analyzer:
•
Set LCD screen contrast
•
Setup Auto-Cal
•
Assign Password
•
Initiate Self -Test
•
Check software version
•
View sensor output
•
Log out.
•
Zero
. Used to set up a zero calibration.
•
Span.
Used to set up a span calibration.
•
Alarms.
Used to set the alarm setpoints and determine whether
each alarm will be active or defeated, HI or LO acting, latching,
and/or failsafe.
•
Range.
Used to set up three analysis ranges that can be switched
automatically with auto-ranging or used as individual fixed
ranges.
Any function can be selected at any time by pressing the appropriate
button (unless password restrictions apply). The order as presented in this
manual is appropriate for an initial setup.
Contrast Function is
DISABLED
(Refer to Section 1.6)
Summary of Contents for 3000PA-EU
Page 2: ......
Page 3: ......
Page 4: ......
Page 56: ...Percent Oxygen Analyzer Operation 4 4 19 Teledyne Analytical Instruments...
Page 77: ...A 10 Appendix Appendix Appendix Appendix Appendix Model 3000PA Teledyne Analytical Instruments...
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......