In run mode the AFA4000 Status Window will show the current status: -
1.
“User present” - the user is detected, the sash drive is inhibited, Tiptronic & manual drive is enabled.
2.
“XX Seconds to close” - the user is not detected; the sash will close following the countdown delay time.
3.
“Sash Closing” will be displayed when the sash is driving closed (auto close, tiptronic, footswitch or
pushbutton).
4.
“Sash Closed” - The user is not present and the sash has closed. (Keypad Backlight dimmed).
5.
“XX Seconds to open” the user is detected; the sash will open following the countdown delay time.(Auto
open enabled)
6.
“Sash Opening” will be displayed when the sash is driving open (auto open, tiptronic, footswitch or
pushbutton).
7.
“Operator Manual Move” will be displayed if the sash is manually opened or closed.
8
.
“Sash Inhibited” will be displayed if the Tilt switch input is open, drive is inhibited until Tilt switch is
closed.
9.
“Sash Disabled” will be displayed if the sash drive has been disabled in the menu settings.
10.
“Sash Locked” will be displayed if the sash is at or above the calibrated lock position.
Auto Sash Operation
and Alarm Indication
If the Auto open feature is enabled in the menu and is switched on from the keypad: “*” (asterix) will be
displayed in the left-hand corner of the Status Window, e.g.
“* User present”.
To enable Auto open press and hold the Mute and
↑
buttons for 1 second or until the “*” symbol is displayed.
To disable Auto open press and hold the Mute and
↑
buttons for 1 second or until the “*” symbol disappears.
The Auto sash can be set to be disabled, enabled or USR enabled, USR enable is User enable / disable.
To User enable the Auto Sash Press the Mute and
↓
buttons for 1 second.
To User disable the Auto Sash Press the Mute and
↓
buttons for 1 second. “USR disabled” will be displayed in
the Status window.