Clear and Sanitize Procedures
7.
Set Serial Port to
Disable
.
8.
Press
F4
and then select
Yes
to save and exit.
Disable the TekLink Port
Complete the following steps to disable the TekLink port on the instrument:
1.
Log on to the instrument as an administrator.
2.
From the Windows Start menu, go to the
Control Panel
.
3.
Select
Hardware and Sound
.
4.
Click
Device Manager
to open the Device Manger window.
5.
Expand the device category Network Adapters.
6.
Double-click the Network Connection.
7.
Click the General tab, and select Device Usage drop-down list.
Remove the Hard Disk
Drive
To remove the hard disk drive assembly, refer to the TLA6400 Series Logic
Analyzer Installation Manual.
All user-storable data is stored on the hard disk drive. Scrub the hard disk drive in
a manner that meets local and any other security requirements that you have.
NOTE.
Before scrubbing the hard disk drive, make certain that you created OS
Restore CD-ROMs. For more information, see Creating Operating System
Restore CD-ROMs in the TLA6400 Series Logic Analyzer Installation Manual.
After the hard disk drive has been scrubbed, reinstall the operating system and
TLA application software.
Scrubbing the hard disk drive will not affect calibration of the instrument. You
can completely erase or remove any secure data without affecting the calibration
of the instrument.
8
TLA6400 Series Declassi
fi