Clear and Sanitize Procedures
Table 2: Nonvolatile memory devices (cont.)
Type and
min. size
Function
Type
of user
info
stored
Method
modi
fi
cation
Data
Input
method
Location
User
accessible
To clear
To sanitize
Flash
2 MByte
(512 X 8)
Stores model
number and
instrument
serial number
None
Indirect
Firmware
operations
Interface
board
No
Load new
fi
rmware.
See installation
manual.
Not applicable,
does not contain
user data
or settings.
Sanitizing
would disable
instrument
functionality.
NVRAM
128 KByte
Stores
instrument
serial number
and calibration
constants.
None
Indirect
Firmware
operations
Acquisition
board
No
Not applicable, does
not contain user
data or settings.
Clearing would
disable instrument
functionality.
Not applicable,
does not contain
user data
or settings.
Sanitizing
would disable
instrument
functionality.
EPROM
8 KByte
Stores
front panel
keyboard code
None
None
Factory
con
fi
guration
Interface
board
No
Not applicable, does
not contain user
data or settings.
Clearing would
disable instrument
functionality.
Not applicable,
does not contain
user data
or settings.
Sanitizing
would disable
instrument
functionality.
PROM
16 Mbit
Contains
FPGA images
for instrument
control
None
None
Factory
con
fi
guration
Interface
board
No
Not applicable, does
not contain user
data or settings.
Clearing would
disable instrument
functionality.
Not applicable,
does not contain
user data
or settings.
Sanitizing
would disable
instrument
functionality.
PROM
4 Mbit
Contains
FPGA image
for instrument
control
None
None
Factory
con
fi
guration
Interface
board
No
Not applicable, does
not contain user
data or settings.
Clearing would
disable instrument
functionality.
Not applicable,
does not contain
user data
or settings.
Sanitizing
would disable
instrument
functionality.
Flash
32 KByte
Stores power
controller code
None
None
Factory
con
fi
guration
Front panel
board
No
Not applicable, does
not contain user
data or settings.
Clearing would
disable instrument
functionality.
Not applicable,
does not contain
user data
or settings.
Sanitizing
would disable
instrument
functionality.
4
TLA6400 Series Declassi
fi
cation and Securities Instructions